A laboratory technician is preparing to quantify a newly synthesized dye
CHEMISTRY · Multi-part · AP · Chemistry · Properties of Substances and Mixtures · Properties of Substances and Mixtures
Problem
A laboratory technician is preparing to quantify a newly synthesized dye. For the chosen analytical wavelength, the dye’s molar absorptivity is 1.87×104L mol−1cm−1. A quartz cuvette with an internal path length of 0.850cm is employed. A solution that is known to be 3.20×10−5M in the dye is placed in the cuvette, and its absorbance is recorded.
Answer the following.
(1) Part 1: (a) Calculate the expected absorbance of the solution. Show your work.
(2) Part 2: (b) The spectrophotometer’s detector is linear only for absorbance values between 0.050 and 1.000. Based on your calculation in part (a), determine whether the measured absorbance will fall within the instrument’s linear range and justify your answer.