A difference amplifier measuring a biological signal is exposed to of common-mode…
PHYSICS · Multi-part · Fermi University Physics · Electronics, Instrumentation & Communication · Feedback, operational amplifiers and analogue signal processing · Sensors, detectors and experimental instrumentation
Electronics for Measurement, Ch The operational amplifier and negative feedback
Problem
A difference amplifier measuring a biological signal is exposed to 15mV of common-mode mains pickup at both inputs. Its own common-mode rejection ratio is CMRR=90dB (a ratio of differential gain to common-mode gain, in decibels using 20log10).
((a)) Convert the 90dB figure to a plain ratio.
((b)) Calculate the residual error this common-mode voltage produces, referred to the input.