A laser-pulse energy meter uses a photodiode of responsivity and capacitance
PHYSICS · Multi-part · Fermi University Physics · Electronics, Instrumentation & Communication · Feedback, operational amplifiers and analogue signal processing · Sensors, detectors and experimental instrumentation
Electronics for Measurement, Ch The operational amplifier and negative feedback
Problem
A laser-pulse energy meter uses a photodiode of responsivity 0.55AW−1 and capacitance 100pF, wired directly into the inverting input of an operational amplifier whose own input capacitance is 5pF and whose gain-bandwidth product is fT=1MHz. The design target is a transimpedance sensitivity of 0.6V per microwatt of optical power, averaged over a pulse.
((a)) Choose the feedback resistor Rf for the design sensitivity, round it to the nearest E12 preferred value, and state the sensitivity it actually gives. Then use the rule Cf≈Cin/(2πRffT), with Cin the total capacitance at the input, to choose the feedback capacitor (again the nearest E12 value) and compute the −3dB bandwidth this leaves the stage.
((b)) The transimpedance output feeds an operational-amplifier integrator (input resistor R1, capacitor C, no resistor across C) that integrates over a laser pulse of width 50μs to give a reading proportional to pulse energy. Choose R1=100kΩ (fixing the integrator's input resistance at that value) and choose C, to the nearest E12 preferred value, so that the time constant R1C equals the pulse width; state the actual time constant achieved and its percentage deviation from 50μs.