A platinum resistance thermometer (RTD) of resistance sits at and is read by a…
PHYSICS · Multi-part · Fermi University Physics · Electronics, Instrumentation & Communication · Feedback, operational amplifiers and analogue signal processing · Sensors, detectors and experimental instrumentation
Electronics for Measurement, Ch Noise and the signal-to-noise ratio
Problem
A platinum resistance thermometer (RTD) of resistance Rs=500Ω sits at 350K and is read by a voltage amplifier with input voltage noise density en=5.0nV/Hz and input current noise density in=2.0pA/Hz. The amplifier's output passes through a first-order low-pass filter of corner frequency fc=200Hz. A constant bias current I=1.0mA is forced through the RTD, whose resistance varies with temperature as dR/dT=0.40ΩK−1 near this operating point. The RTD's Johnson noise, the amplifier's voltage noise and the voltage the amplifier's current noise produces across Rs are independent and are all referred to the amplifier's input.
((a)) Find the three noise voltage densities referred to the input, combine them by a quadrature sum, and state which one dominates the noise power and by what percentage.
((b)) Find the noise-equivalent bandwidth of the first-order low-pass filter.
((c)) Find the rms noise voltage referred to the input.
((d)) Find the smallest temperature change this instrument can detect at a signal-to-noise ratio of one.