A decoupling stage behaves as an RC low-pass with

PHYSICS · Multi-part · Fermi University Physics · Electronics, Instrumentation & Communication · Signals, systems, control and digital signal processing · Signals, systems, control and digital signal processing

Electronics for Measurement, Ch Linear systems, the impulse response and convolution

Problem

A decoupling stage behaves as an RC low-pass with RC=200 μsRC = 200\,\mu\mathrm{s}. A technician injects a rectangular test pulse of area 5.0 μV s5.0\,\mu\mathrm{V\,s} and width 500 μs500\,\mu\mathrm{s}, and measures a peak output of 9.2 mV9.2\,\mathrm{mV}. They argue: "The ideal impulse response should peak at h(0)=1/RC=5000 s−1h(0) = 1/RC = 5000\,\mathrm{s^{-1}}, so for a pulse of area 5.0 μV s5.0\,\mu\mathrm{V\,s} the peak should be (5.0 μV s)(5000 s−1)=25 mV(5.0\,\mu\mathrm{V\,s})(5000\,\mathrm{s^{-1}}) = 25\,\mathrm{mV}. I measured only 9.2 mV9.2\,\mathrm{mV}, which is 63%63\% low. Since I trust the pulse generator's area, the stage's true time constant must be about 63%63\% longer than its nameplate value, i.e. RC≈326 μsRC \approx 326\,\mu\mathrm{s}, not 200 μs200\,\mu\mathrm{s}." ((a)) Find the error in this reasoning. ((b)) Give the correct interpretation of the 9.2 mV9.2\,\mathrm{mV} reading, and show it is consistent with RC=200 μsRC = 200\,\mu\mathrm{s}.