A calibration engineer writes the following justification for skipping any check of…
PHYSICS · Multi-part · Fermi University Physics · Electronics, Instrumentation & Communication · Feedback, operational amplifiers and analogue signal processing · Sensors, detectors and experimental instrumentation
Electronics for Measurement, Ch Sensors, transducers and the measurement front end
Problem
A calibration engineer writes the following justification for skipping any check of the interface wiring: The 350Ω quarter-bridge strain gauge is fully specified by its sensitivity of 2.5μV per microstrain at 5V excitation. Once I know that number I can multiply it by the amplifier gain and I am done. Using this reasoning, for a measured strain of 240με and a downstream difference amplifier of gain 1000, the engineer predicts an amplifier output of 240×2.5μV×1000=600mV. The difference amplifier has a differential input resistance of 2.1kΩ, and the bridge (all four arms equal at 350Ω) presents a source resistance of 350Ω at its output terminals.
((a)) Name the entry missing from the engineer's specification of the sensor and explain why it changes the answer here.
((b)) Calculate the amplifier's actual output, and state by how many millivolts and what percentage the engineer's prediction is wrong.