A diagnostic ultrasound scanner can be operated at 2.0 MHz or at 8.0 MHz
PHYSICS · Multi-part · Fermi University Physics · Oscillations, Waves & Acoustics · Acoustics of fluids: generation, propagation and attenuation · Acoustics of fluids: generation, propagation and attenuation
Fluids and Continuum Physics, Ch Acoustics: sound propagation, radiation and attenuation
Problem
A diagnostic ultrasound scanner can be operated at 2.0 MHz or at 8.0 MHz. To estimate how the two settings differ in penetration, model the tissue as water and use the classical absorption coefficient, which accounts for viscosity alone, α=2ω2μ/(3ρ0c3), with the amplitude of a plane wave falling as e−αx. For water take μ=1.0×10−3 Pa s, ρ0=1000 kg m−3 and c=1483 m s−1.
((a)) Estimate α at 2.0 MHz.
((b)) By what factor is α larger at 8.0 MHz than at 2.0 MHz?
((c)) Estimate the distance over which the amplitude falls to 1/e of its starting value at 8.0 MHz.