A cryo-electron microscope operates at

PHYSICS · Subjective · Fermi University Physics · Optics & Photonics · Biophysics & Medical Physics · Macromolecules, membranes and cell architecture

Forces, Machines and Signals, Ch Seeing molecules: diffraction, cryo-electron microscopy and fluorescence

Problem

A cryo-electron microscope operates at 300 kV300\,\mathrm{kV}, giving electrons of wavelength λ=1.97 pm\lambda = 1.97\,\mathrm{pm}. A micrograph is recorded with defocus Δf=0.50 μm\Delta f = 0.50\,\mathrm{\mu m}. Using the first-zero condition q=1/λΔfq = 1/\sqrt{\lambda\Delta f} for the contrast transfer function (ignoring spherical aberration), find the real-space spacing d=1/qd = 1/q corresponding to the first CTF zero.