A silica bead of radius is held in a single-beam optical trap in water of viscosity at ( )
PHYSICS · Multi-part · Fermi University Physics · Biophysics & Medical Physics · Molecular machines, transport and bioenergetics · Molecular machines, transport and bioenergetics
Forces, Machines and Signals, Ch Measuring one molecule: traps, tweezers and single-molecule fluorescence
Problem
A silica bead of radius a=0.50μm is held in a single-beam optical trap in water of viscosity η=1.0×10−3Pas at T=298K (kBT=4.11pNnm), with a measured stiffness κ=0.050pNnm−1. Treat the bead as a sphere in an unbounded fluid, so its drag coefficient is ζ=6πηa, and treat the trap as a harmonic spring.
((a)) The bead's position x(t) obeys the overdamped Langevin equation ζx˙=−κx+ξ(t), with a random thermal force satisfying ⟨ξ(t)ξ(t′)⟩=2ζkBTδ(t−t′). Starting from this equation, derive the one-sided power spectral density Sx(f) and identify the corner frequency fc in terms of κ and ζ. Compute fc for this bead, and show, by integrating Sx(f) over all frequencies, that the result recovers the equipartition value of ⟨x2⟩.
((b)) A single bond is pulled by a force ramp f=rt. Its instantaneous rupture rate is k(f)=k0exp(fx‡/kBT), where k0 is the unstressed off-rate and x‡ the distance to the transition state along the pulling direction. Starting from the survival probability S(t) obeying dS/dt=−k(rt)S, derive the most probable rupture force f∗ (the mode of the rupture-force distribution) as a function of r. Evaluate f∗ for a receptor-ligand bond with k0=0.010s−1, x‡=0.30nm, pulled at r=1.0×103pNs−1, at T=298K.