PHYSICS · Multi-part · Fermi University Physics · Fluids & Continuum Physics · Biophysics & Medical Physics · Bioelectricity, the senses and the physics of the body
Forces, Machines and Signals, Ch Bones, blood and body size
Problem
A renal artery is modelled as a thin-walled, incompressible, inviscid elastic tube of radius r=2.5mm, wall thickness hw=1.0mm, carrying blood of density ρ=1060kgm−3. Here hw/r=0.4 is not small, so the thin-wall results below give an estimate, not an exact speed. The wall obeys Hooke's law with Young's modulus E, and wave reflections are ignored.
((a)) From the hoop stress σθ=pr/hw and Hookean strain Δr/r=Δσθ/E, derive dA/dp=2πr3/(Ehw) for the cross-sectional area A=πr2.
((b)) Starting from linearised continuity, ∂A/∂t+A0∂u/∂x=0, and momentum, ρ∂u/∂t=−∂p/∂x, combine them into a wave equation for p(x,t) and show that the pulse-wave speed is cpw=Ehw/2ρr.
((c)) In a healthy young wall, E=0.50MPa. Evaluate cpw.
((d)) With age the wall stiffens to E=2.0MPa, with hw and r unchanged. Find the new cpw and state, in words, how cpw scales with E.