A 50 m signal run must be installed between a fast pulse source and a coincidence…

PHYSICS · Multi-part · Fermi University Physics · Electricity, Magnetism & Electromagnetism · Electromagnetic waves and their propagation · Electromagnetic waves and their propagation

Inside the Instrument, Ch Transmission lines, grounding and the bench

Problem

A 50 m signal run must be installed between a fast pulse source and a coincidence unit; both ends are matched to 50 Ω50\,\Omega, so reflections are not an issue here. The run must carry a 100 MHz100\,\mathrm{MHz} calibration burst with no more than 4 dB4\,\mathrm{dB} of insertion loss, and isolated logic edges, each arriving after the line has settled from the one before, which must reach at least 90 %90\,\% of their final value within 20 ns20\,\mathrm{ns} of arriving. Two cables are available: Belden 8240 (RG-58), specified at 3.8 dB3.8\,\mathrm{dB} per 100 ft100\,\mathrm{ft} at 100 MHz100\,\mathrm{MHz}, for which the chapter's matched-line model of a 30 m30\,\mathrm{m} length reaches 90 %90\,\% of its step height about 19 ns19\,\mathrm{ns} after the step arrives. Belden 9913, specified at 1.5 dB1.5\,\mathrm{dB} per 100 ft100\,\mathrm{ft} at 100 MHz100\,\mathrm{MHz}. Use 1 ft=0.3048 m1\,\mathrm{ft} = 0.3048\,\mathrm{m}. Assume both cables follow the chapter's pure f\sqrt f loss model, each fixed by its own 100 MHz100\,\mathrm{MHz} figure. ((a)) Find the total insertion loss, in decibels, of a 50 m50\,\mathrm{m} run of each cable at 100 MHz100\,\mathrm{MHz}, and state which (if either) meets the 4 dB4\,\mathrm{dB} budget. ((b)) From the model's matched step response, find how the 90 %90\,\% settling time depends on a cable's 100 MHz100\,\mathrm{MHz} loss over the run. Use it with the 30 m30\,\mathrm{m} figure for Belden 8240 to estimate the 90 %90\,\% settling time of a 50 m50\,\mathrm{m} run of each cable, and state whether each meets the 20 ns20\,\mathrm{ns} requirement. ((c)) Recommend a cable for the run, giving the physical reason, in terms of R′R', why a lower-loss cable is also expected to settle faster for the same length.

Browse the problem library