A signal-conditioning group must choose between a bipolar-input and a MOS-input…
PHYSICS · Multi-part · Fermi University Physics · Electronics, Instrumentation & Communication · Feedback, operational amplifiers and analogue signal processing · Sensors, detectors and experimental instrumentation
Inside the Instrument, Ch The differential pair and the current mirror
Problem
A signal-conditioning group must choose between a bipolar-input and a MOS-input operational amplifier for two different transducers, using the typical input bias currents IB,bip=80nA (as for the 741) and IB,MOS=2pA (typical of a CMOS input pair).
((a)) A glass pH electrode presents a source resistance Rs1=1.0×109Ω and its output changes by 59mV per pH unit. The design requires the voltage error caused by input bias current to correspond to no more than 0.1 pH unit. Compute the bias-current-induced voltage error for each amplifier type at this source resistance and state whether each meets the specification.
((b)) A strain-gauge bridge with 350Ω arms presents a source resistance of at most Rs2=350Ω to each input and delivers a signal of order 1mV. Compute the bias-current-induced voltage error for each amplifier type at this source resistance, and compare each to the 1mV signal.
((c)) State, with a reason tied to a number from (a) or (b), which amplifier type should be used for each transducer, and explain in one sentence why bias current alone does not decide the choice for the bridge.