PHYSICS · Multi-part · Fermi University Physics · Scientific Practice & Experimental Methods · Biophysics & Medical Physics · Uncertainty in measurement
Life at Thermal Energy, Ch Echoes, precession and the therapeutic beam
Problem
A sonographer scans a carotid artery with a 4MHz probe. The returned Doppler shift is measured very precisely as Δf=150Hz. The beam-to-flow angle cursor is set to θ=87∘, but the operator can only align the cursor with the vessel wall to within ±3∘. Take the speed of sound in soft tissue as cs=1540ms−1.
((a)) Using θ=87∘, calculate the blood speed v from Δf=2fvcosθ/cs.
((b)) Recalculate v for the two extreme angles the ±3∘ uncertainty allows, θ=84∘ and θ=90∘.
((c)) The sonographer reports the answer from part (a) as a reliable measurement, reasoning that Δf itself was measured to three significant figures. Explain why this reasoning is wrong.