Problem

An open MRI scanner uses a field of B0=1.0 TB_0 = 1.0\,\mathrm{T}, and the patient is at body temperature, T=310 KT = 310\,\mathrm{K} (kBT=4.28×10−21 Jk_BT = 4.28\times10^{-21}\,\mathrm{J}). Take γ/2π=42.58 MHz T−1\gamma/2\pi = 42.58\,\mathrm{MHz\,T^{-1}} and h=6.63×10−34 J sh = 6.63\times10^{-34}\,\mathrm{J\,s}. ((a)) Calculate the proton resonance frequency f0f_0 and the level gap ΔE=hf0\Delta E = hf_0. ((b)) Starting from the Boltzmann factor e−ΔE/kBTe^{-\Delta E/k_BT} for the two spin populations and using e−x≈1−xe^{-x}\approx 1-x for small xx, derive the fractional population excess Δn/n≈ΔE/2kBT\Delta n/n \approx \Delta E/2k_BT, and evaluate it at 1.0 T1.0\,\mathrm{T}. ((c)) A 1 mm31\,\mathrm{mm^3} voxel of water contains 6.7×10196.7\times10^{19} protons. How many more protons align with the field than against it?