A digital multimeter on its current range resolves and is specified to
PHYSICS · Multi-part · Fermi University Physics · Scientific Practice & Experimental Methods · Quantities, units and the practice of measurement · Uncertainty in measurement
Measurement and Method, Ch The instrument: resolution, sensitivity, calibration and traceability
Problem
A digital multimeter on its 200.00mA current range resolves 0.01mA and is specified to ±(1.0%ofreading+3digits), where one digit is 0.01mA on this range. Evaluate the bound on the reading's error, in milliamps and as a percentage of the reading, for displayed readings of:
((a)) 180.00mA
((b)) 20.00mA
((c)) 2.00mA