Two marks on a laboratory track are read off the same steel rule
PHYSICS · Multi-part · Fermi University Physics · Scientific Practice & Experimental Methods · Uncertainty in measurement · Uncertainty in measurement
Measurement and Method, Ch Propagating uncertainty into the quantity you actually want
Problem
Two marks on a laboratory track are read off the same steel rule, at positions x1=85.00cm and x2=32.00cm. The rule's own length may be systematically long or short by a fraction κ, with u(κ)=0.15 per cent (one standard deviation), and this scale error shifts every reading taken with the rule by the same fraction. Apart from this shared scale error, the two readings are exact. The separation of the marks is Δx=x1−x2. Someone argues: "x1 and x2 share the same rule, so they are correlated, and correlation always makes an uncertainty bigger than treating the readings as independent. Treating them as independent gives u(x1)=0.15%×85.00cm=0.128cm and u(x2)=0.15%×32.00cm=0.048cm, so u(Δx)=0.1282+0.0482cm=0.14cm; the correlated value must be larger than this."
((a)) Write x1 and x2 in terms of the true rule readings ℓ1, ℓ2 and the shared scale error κ, and use this to find the correct u(Δx).
((b)) Say whether the quoted claim is right, and explain the error in the reasoning.