A digital caliper with a resolution (display step) of is used to measure the diameter…

PHYSICS · Multi-part · Fermi University Physics · Scientific Practice & Experimental Methods · Uncertainty in measurement · Uncertainty in measurement

Measurement and Method, Ch Type A, Type B and the uncertainty budget

Problem

A digital caliper with a resolution (display step) of 0.01 mm0.01\,\mathrm{mm} is used to measure the diameter of a steel rod eight times, giving the readings (mm): 12.02, 12.05, 12.03, 12.04, 12.06, 12.03, 12.05, 12.04. ((a)) Evaluate the Type A standard uncertainty of the mean diameter, stating its degrees of freedom. ((b)) The manufacturer states only the display resolution as a limit on reading a single value, with no reason to prefer any part of the step. Treat this as a rectangular distribution and evaluate the Type B standard uncertainty it contributes to a single reading.