Problem
A wire's diameter is measured with a micrometer. Four contributions to the uncertainty budget have been evaluated: the scatter of five repeated closures on the wire, u=0.0021mm, evaluated from the standard deviation of the readings (Type A); the zero-point (jaw-closure) correction, estimated from twenty repeated closures on the jaws alone with nothing between them, u=0.0015mm, again from the scatter of repeated readings (Type A); the manufacturer's specification, a bound of ±0.005mm on any single reading, treated as rectangular (Type B), giving u=0.005/3=0.0029mm; a bound of ±0.001mm on thermal expansion of the wire between calibration and measurement, treated as rectangular (Type B), giving u=0.001/3=0.00058mm. All four sensitivity coefficients are 1. A laboratory report combines them as follows: 'Type A means a random effect and Type B means a systematic one, so they cannot be mixed. The Type A total is uA=0.00212+0.00152=0.0026mm and the Type B total is uB=0.00292+0.000582=0.0030mm. The combined standard uncertainty is their sum, uc=uA+uB=0.0056mm.'
((a)) State the error in this reasoning.
((b)) Give the correct combined standard uncertainty uc, to two significant figures.