PHYSICS · Subjective · Fermi University Physics · Scientific Practice & Experimental Methods · Uncertainty in measurement · Uncertainty in measurement
Measurement and Method, Ch Type A, Type B and the uncertainty budget
Problem
A resistor's resistance is found from Ohm's law, R=V/I. The ammeter used is believed to have a small zero-offset δoff on its current reading, with best estimate zero, so the measurement model is R=V/(I−δoff). Eight independent voltmeter readings across the resistor give a mean Vˉ=5.000V with sample standard deviation s=0.006V. The voltmeter's calibration certificate states the reading is correct to within ±0.5% of reading at coverage factor k=2. The ammeter reads I=0.2000A (a single reading, judged stable) with display resolution 0.0001A. The zero-offset δoff is judged, from repeated checks with the circuit open, to lie within ±0.0002A of zero, with nothing favouring any part of that range. (a) Evaluate the standard uncertainty of each row (state the type and distribution assumed for each). (b) Combine the two voltage rows into u(V) and the two current rows into u(I). (c) Propagate to uc(R) and quote R with its standard uncertainty. (d) Identify the dominant row and say what should be improved first.