A thermocouple's output emf is recorded at eleven junction temperatures set with an…
PHYSICS · Multi-part · Fermi University Physics · Scientific Practice & Experimental Methods · Data analysis and model fitting in the laboratory · Data analysis and model fitting in the laboratory
Measurement and Method, Ch Graphs, straight lines and least squares
Problem
A thermocouple's output emf ε is recorded at eleven junction temperatures θ set with an oil bath whose reference thermometer has standard uncertainty u(θ)=0.5K. Repeated readings give a vertical uncertainty u(ε)=2μV at each temperature. The calibration slope is 40μVK−1. A least-squares line through the eleven points, with every point given the 2μV vertical uncertainty alone, gives χ2=200.
((a)) Calculate the horizontal-axis contribution to each point's error bar from the bath's temperature uncertainty, in microvolts, and state whether it may be neglected next to the 2μV vertical bar.
((b)) The fit has ν=9. Calculate χν2 as the fit reports it.
((c)) Recompute χν2 with each point's full uncertainty, the vertical bar and the horizontal contribution of part (a) combined in quadrature, and say whether the line deserves belief.