A DC amplifier used to measure a slowly varying voltage has a noise spectral density

PHYSICS · Multi-part · Fermi University Physics · Scientific Practice & Experimental Methods · Uncertainty in measurement · Designing and running experiments

Measurement and Method, Ch Designing the measurement: noise, averaging, null and differential methods

Problem

A DC amplifier used to measure a slowly varying voltage has a noise spectral density en=50 nV/Hze_n = 50\,\mathrm{nV/\sqrt{Hz}}. Before its case has settled to a steady temperature its zero drifts at a rate D=0.20 μV s−1D = 0.20\,\mathrm{\mu V\,s^{-1}}. Each measurement is the difference between two averages, each of readings taken for a time tt, and the total standard uncertainty on that difference is u(t)=en2t+(Dt)2.u(t) = \sqrt{\frac{e_n^2}{t} + (Dt)^2}. ((a)) Find the optimum averaging time toptt_{\rm opt} and the resulting minimum uncertainty u(topt)u(t_{\rm opt}). ((b)) A thermally regulated enclosure is fitted, cutting the drift by a factor of 88 to D′=25 nV s−1D' = 25\,\mathrm{nV\,s^{-1}}. Find the new optimum averaging time topt′t_{\rm opt}' and the new minimum uncertainty u(topt′)u(t_{\rm opt}'). ((c)) State, as a ratio, how much longer the new averaging takes and how much the minimum uncertainty improves, and comment on whether these two ratios are equal.