Two claims are made about two different apparatus

PHYSICS · Multi-part · Fermi University Physics · Scientific Practice & Experimental Methods · Electronics, Instrumentation & Communication · Feedback, operational amplifiers and analogue signal processing

Measurement and Method, Ch Designing the measurement: noise, averaging, null and differential methods

Problem

Two claims are made about two different apparatus. Neither is correct as stated. ((a)) A demonstrator averages a step of size 0.50 μV0.50\,\mathrm{\mu V} buried in noise that, for an averaging time tt, falls as 0.080/t μV0.080/\sqrt{t}\,\mathrm{\mu V} (with tt in seconds), while a slow drift moves the baseline by DtD t with D=0.020 μV s−1D = 0.020\,\mathrm{\mu V\,s^{-1}}. A quick calculation gives an optimum averaging time of a few seconds, but the demonstrator argues: "averaging always helps, so I will average for 10 s10\,\mathrm{s}, five times longer than that estimate, to get the best possible signal-to-noise ratio." Find the true optimum averaging time toptt_{\rm opt} and the resulting S/NS/N, and compare it with the S/NS/N the demonstrator actually gets at 10 s10\,\mathrm{s}. ((b)) A strain-gauge bridge has four arms each of nominal resistance R=120 ΩR = 120\,\mathrm{\Omega}, supplied at V=5.0 VV = 5.0\,\mathrm{V}, and is balanced against a decade resistor before loading. A strain changes one arm's resistance by ΔR=1.0×10−3 Ω\Delta R = 1.0\times10^{-3}\,\mathrm{\Omega}, unbalancing the bridge by ΔVout=V ΔR/(4R)\Delta V_{\rm out} = V\,\Delta R/(4R). Someone argues: "this measurement can only be trusted if the null detector is a voltmeter with a calibration accuracy of 1 %1\,\%, because accuracy is what determines how well we know the voltage." Compute ΔVout\Delta V_{\rm out} and explain what is wrong with this argument, stating what property of the detector the measurement really needs.

Browse the problem library