The illumination channel of an endoscope uses a light-source objective of numerical…
PHYSICS · Multi-part · Fermi University Physics · Optics & Photonics · Wave propagation and light in media · Beams, resonators and guided-wave photonics
Optics and Photonics, Ch Optical fibres in communication and endoscopy
Problem
The illumination channel of an endoscope uses a light-source objective of numerical aperture NAobj=0.18 (Chapter 2 definition) to launch light into the unordered illumination fibre bundle. Each fibre in the bundle has NAfibre=0.24 and core index n1=1.50. The illumination cable is L=8 km long (a long feed cable in a laboratory test rig) with absorption coefficient α=1.15×10−5m−1 (Chapter 5 definition). Neither end face is coated.
((a)) Assuming the objective delivers its power uniformly over its solid angle (small-angle approximation), what fraction of the launched power falls within a fibre's acceptance cone?
((b)) Find the total loss budget in decibels, combining the normal-incidence Fresnel reflectance (Chapter 6) at both end faces with the absorption over the full 8 km.