A calorimeter combines two independently measured temperature differences into an…
PHYSICS · Multi-part · Fermi University Physics · Scientific Practice & Experimental Methods · Uncertainty in measurement · Uncertainty in measurement
Scientific Computing for Physicists, Ch The uncertainty of a simulated number: inputs, sensitivity and correlated output
Problem
A calorimeter combines two independently measured temperature differences into an energy estimate y=c1x1+c2x2, with fixed calibration coefficients c1=3.0JK−1 and c2=−2.0JK−1. The readings are x1=12.0±0.5K and x2=8.0±1.0K, independent, each evaluated Type A from repeated readings of the probe.
((a)) Find y and its linearly propagated uncertainty uy.
((b)) Compute the first-order sensitivity index of each input.
((c)) State which reading should be measured more precisely to reduce uy most.