A radiographer argues: "Since D=ΦE(μen/ρ)D = \Phi E (\mu_{\rm en}/\rho) everywhere in a photon beam, a detector placed exactly on the skin surface of a patient undergoing a 6 MV6\,\mathrm{MV} therapy beam reads the same dose as the kerma computed at that point from the incident fluence and the coefficients of the material." For this beam the secondary electrons set in motion by the photons have ranges of up to about 3 cm3\,\mathrm{cm}, and the depth-dose curve reaches its maximum, where dose and kerma track each other closely, at about 1.5 cm1.5\,\mathrm{cm} depth. ((a)) Is the radiographer's claim correct at the skin surface? Identify the geometrical condition that is violated there. ((b)) State, with a reason, whether the true surface dose is greater than, equal to, or less than the kerma computed at the surface.