A laboratory fibre-fed spectrometer uses a step-index fibre of core diameter and…
PHYSICS · Multi-part · Fermi University Physics · Optics & Photonics · Geometrical optics and imaging · Photodetection, photonic sources and quantum optics
What an Instrument Can See, Ch Radiance, etendue and a throughput budget that is spent
Problem
A laboratory fibre-fed spectrometer uses a step-index fibre of core diameter 100μm and numerical aperture NA=0.10, coupled losslessly into free space.
((a)) Treating the fibre exit face as a circular Lambertian aperture, find its etendue Ξ=πANA2, where A is the face area.
((b)) The detector records a photoelectron signal of 50e−s−1 over an integration time of 20s, with dark current 0.02e−s−1 and read noise 3e− (rms, added once per read, independent of exposure time). Find the signal-to-noise ratio and state which noise term dominates.
((c)) A 650nm laser diode of optical power 2.0mW is coupled into the fibre. Using the photopic luminous efficiency V(650nm)=0.107 and the peak luminous efficacy Km=683lmW−1, find the luminous flux this power corresponds to, and state why this conversion is only valid for this one wavelength.