A fibre-fed echelle spectrograph for a telescope must reach a resolving power with a…
PHYSICS · Multi-part · Fermi University Physics · Optics & Photonics · Geometrical optics and imaging · Geometrical optics and imaging
What an Instrument Can See, Ch Radiance, etendue and a throughput budget that is spent
Problem
A fibre-fed echelle spectrograph for a 4.0m telescope must reach a resolving power R=60000 with a fibre that subtends 1.5 arcseconds on the sky. Its echelle has tanθbl=2.0. Assume the fibre carries Dtelϕ to the collimator without growth. Take the seeing as Gaussian with a full width at half maximum of 1.0 arcsecond, so that a centred fibre of angular diameter d, in arcseconds, catches a fraction 1−2−d2 of a star's light.
((a)) Find the collimated beam diameter Dcoll that reaches R=60000 with the unsliced fibre.
((b)) The beam can be at most 200mm across. An image slicer that cuts the fibre image into N equal strips and stacks them along the slit divides the width the collimator sees by N. Find the smallest N that reaches R=60000.
((c)) Instead of slicing, the fibre could be narrowed, with the 200mm beam. Find the angular diameter that reaches R=60000, and the fractions of a star's light caught by that fibre and by the 1.5 arcsecond fibre.