A pure phase specimen has amplitude transmittance with

PHYSICS · Multi-part · Fermi University Physics · Optics & Photonics · Fourier optics and image formation · Fourier optics and image formation

What an Instrument Can See, Ch Seeing what the transfer function forbids: phase contrast and super-resolution

Problem

A pure phase specimen has amplitude transmittance t(x)=eiϕ(x)t(x) = e^{i\phi(x)} with ∣ϕ(x)∣≪1|\phi(x)| \ll 1. It is illuminated coherently by a structured pattern Uill(x)=1+cos⁡(2πνix)U_{\mathrm{ill}}(x) = 1 + \cos(2\pi\nu_i x) of the kind used in structured illumination, with νi\nu_i chosen at 0.80.8 times the incoherent cutoff of the objective. The transmitted field is U(x)=Uill(x) t(x)U(x) = U_{\mathrm{ill}}(x)\,t(x), and this field passes, unattenuated at every spatial frequency present, through an aberration-free lens acting as a Fourier transformer (with the usual transfer function of an imaging system) onto a camera that records I(x)=∣U(x)∣2I(x) = |U(x)|^2. ((a)) Expand U(x)U(x) to first order in ϕ(x)\phi(x) and show that the terms proportional to ϕ(x)\phi(x) in I(x)I(x) vanish, for any real illumination profile including this structured one. ((b)) For a phase amplitude ϕ0=0.06\phi_0 = 0.06 rad, keep the next term, eiϕ≈1+iϕ−ϕ2/2e^{i\phi}\approx1+i\phi-\phi^2/2, and find the phase-dependent irradiance contrast to second order. Compare it with the value for uniform illumination at the same ϕ0\phi_0. ((c)) In one sentence, explain why the resolution gain of structured illumination does not rescue a pure phase object.