Ion Chromatography and Suppression
A municipal water lab needs to report nitrate, sulfate, chloride, and fluoride in a single ten-minute run. It must do this at parts-per-billion levels. The sample also contains a thousand-fold excess of sodium and carbonate eluent. This is the daily reality that ion chromatography (IC) with suppressed conductivity detection was built to solve. Unlike UV or fluorescence detection, conductivity responds to every ion in the eluent stream. This is both the appeal and the problem. The background electrolyte needed to elute ions off an Ion-Exchange Chromatography column is itself enormously more conductive than the trace analytes riding through it. Suppression is the trick that makes conductivity detection selective and sensitive at once. It chemically converts the high-conductivity eluent into a low-conductivity form immediately before the detector. Simultaneously, it converts the analyte ions into their high-conductivity acid or base forms. The background drops by two to three orders of magnitude. The analyte signal is amplified. A technique that looks impossible on paper becomes the workhorse method for anions and cations. It is used in environmental, pharmaceutical, semiconductor, and clinical laboratories. The physical picture is deceptively simple. A membrane or a packed resin swaps counter-ions with the eluent stream. But the electrochemistry rewards careful study. So do the equilibria that govern residual background conductivity. So does the quantitative relationship between detector signal and concentration. Suppression is a continuous acid–base and ion-exchange process operating under flow. Understanding it means understanding conductance itself. It means understanding how ionic mobility, concentration, and charge combine to produce a measurable signal. It also means understanding how a Donnan-excluded membrane can run for months. Such a membrane can exchange millions of ions without ever needing regeneration by hand. This article works through the physics of conductivity detection. It covers the two dominant suppressor architectures: membrane-based continuous suppression and disposable/regenerable packed-bed suppression. It explains the equilibrium and kinetic reasons suppression works as well as it does. It also presents three fully quantitative worked examples. These range from computing raw background conductance, to designing a calibration for sulfate at low ppb, to diagnosing a "ghost peak" caused by suppressor capacity exhaustion.
- Conductivity detection sees every ion, so a high-conductivity eluent (needed for Ion-Exchange Chromatography elution) would swamp trace analyte signals without further treatment.
- A suppressor continuously exchanges the eluent's co-ion for water (converting NaHCO₃/Na₂CO₃ eluent to weakly-dissociated H₂CO₃) while converting analytes to their strong acid/base forms.
- Suppression cuts background conductance by 10²–10³× and simultaneously boosts analyte conductance by converting, e.g., NaCl to HCl (H⁺ mobility ≈ 5× Na⁺ mobility).
- Membrane suppressors (electrolytic, continuously regenerated) dominate modern practice; packed-bed suppressors are cheaper but must be periodically regenerated or replaced.
- Detector response is G = \kappa \cdot (A/L) \cdot 10^{6} in µS, with built from each ion's equivalent conductance and concentration — the Kohlrausch law of independent ionic migration.
- Key idea: suppression converts a conductivity detector from a universal, low-selectivity sensor into a high-sensitivity, low-background instrument by simultaneously removing the eluent's charge carriers and amplifying the analyte's.
Where this lives today
Suppressed conductivity IC is the default method (EPA Method 300.0/300.1 and ISO 10304 series) for anions in drinking water, wastewater, and environmental monitoring — chloride, nitrate, nitrite, sulfate, fluoride, phosphate, and bromate are routinely quantified in a single 10–20 minute run at low-ppb detection limits. Semiconductor fabs use suppressed IC to monitor ultrapure water for trace anionic contaminants (down to sub-ppb) that could poison wafer processing. Pharmaceutical quality control uses cation-suppressed IC to quantify counter-ions (Na⁺, K⁺, Ca²⁺) in drug formulations and to check for residual catalysts or excipient contaminants. Clinical and biomedical laboratories use suppressed IC for anion panels in biological fluids, and industrial hygiene labs use it to quantify oxyhalides (bromate, chlorite, chlorate) in disinfected water, a use case that depends critically on the low, stable suppressed background achievable only because the suppressor continuously converts a strongly conductive hydroxide or carbonate eluent into water or a very weak acid, preserving the sensitivity that makes trace-level oxyhalide monitoring possible at all. The same core hyphenation strategy — using an electrochemical or conductivity detector downstream of an Ion-Exchange Chromatography separation — underlies related techniques like capillary electrophoresis with conductivity detection, extending the suppression logic beyond pressure-driven liquid chromatography entirely.
Ask the tutor: How would the background conductivity and suppression gain change if the eluent were switched from carbonate/bicarbonate to a dilute KOH gradient generated electrolytically from water?
The physical picture
A conductivity detector measures the ability of the mobile phase to carry electric current between two electrodes, expressed as conductance (siemens, S) or as specific conductance (S/cm, more often µS/cm for aqueous work). Every ion in solution — eluent counter-ion, eluent co-ion, and analyte alike — contributes to according to the Kohlrausch law of independent migration: each ionic species contributes a term proportional to its concentration, its charge, and its own limiting equivalent conductance , essentially independent of what other ions are present (at the low ionic strengths typical of IC eluents, before ion-pairing or activity corrections become significant). The trouble is immediate: a carbonate/bicarbonate eluent used to elute anions like Cl⁻, NO₃⁻, and SO₄²⁻ from a high-capacity anion exchanger might run at 4.5 mM Na₂CO₃ / 1.4 mM NaHCO₃, giving a background conductivity in the hundreds of µS/cm — while the analyte anions of interest sit at 0.05–5 mg/L, contributing only fractions of a µS/cm on top of that background. Detecting a small signal riding on a huge, noisy baseline is the classic small-signal-on-large-background problem, and simply subtracting a baseline does not recover the lost signal-to-noise ratio, because baseline noise (pump pulsation, temperature drift) scales with the baseline itself. Suppression solves this by chemical conversion immediately downstream of the column, before the detector cell. For anion analysis, the suppressor is a strong-acid cation exchanger in the H⁺ form. As the eluent stream passes through, Na⁺ (from Na₂CO₃/NaHCO₃) is exchanged for H⁺:
Carbonic acid is a weak acid () and barely dissociates at the resulting concentrations, so the eluent's contribution to conductivity collapses. Simultaneously, every analyte anion — say Cl⁻ paired with Na⁺ in solution — is converted to its strong acid form:
HCl is a strong acid, fully dissociated, and H⁺ has an enormous limiting equivalent conductance () compared to Na⁺ (). The analyte signal is amplified roughly 7-fold in the cation contribution alone, while the background is suppressed 100- to 1000-fold. For cation analysis (using an acidic eluent like methanesulfonic acid, MSA), the suppressor is a strong-base anion exchanger in the OH⁻ form, converting MSA to water and analyte cations (e.g., Na⁺ paired with MSA⁻) to strong bases (NaOH).
Modern suppressors are almost universally membrane-based and electrolytically regenerated (e.g., Dionex's Self-Regenerating Suppressor, or Thermo Fisher's/Metrohm's equivalents), meaning the H⁺ (or OH⁻) needed for exchange is generated in situ by electrolysis of water, using an external regenerant reservoir of only deionized water and a small applied current — so the suppressor never needs manual chemical recharging and can run continuously for months.
The core ideas
The detector's raw signal is specific conductance, related to measured conductance by the cell constant:
where:
- = specific conductance (S/cm), usually reported in µS/cm
- = measured conductance (S), the reciprocal of measured resistance
- = the cell constant (cm⁻¹), a fixed geometric property of the electrode pair (typical IC cells: –, calibrated with a standard KCl solution)
Reading: conductance is what the electronics actually measure; specific conductance is the intrinsic solution property, recovered by multiplying by the (fixed, pre-calibrated) cell constant. Everything else in this article is really about predicting from solution composition. Kohlrausch's law of independent ionic migration gives as a sum over all ionic species:
where:
- = limiting (infinite-dilution) equivalent (molar, per unit charge) conductance of ion , in S·cm²/mol (values in Table 1)
- = magnitude of the ionic charge (so the sum is really over equivalents, not moles, for multiply-charged ions)
- = molar concentration of ion (mol/L)
- the factor converts mol/L·(S·cm²/mol) into S/cm units consistently (since 1 L = 1000 cm³)
Reading: every ion adds its own independent term; more mobile ions (larger ) and higher charge contribute disproportionately. This is exactly why swapping Na⁺ (50.1) for H⁺ (349.8) on the analyte, while swapping Na⁺ for H⁺ on a weak-acid anion (which then barely ionizes) on the eluent side, produces such a large net gain in signal-to-background.
| Ion | (S·cm²/mol) | Ion | (S·cm²/mol) |
|---|---|---|---|
| H⁺ | 349.8 | OH⁻ | 199.1 |
| Na⁺ | 50.1 | Cl⁻ | 76.3 |
| K⁺ | 73.5 | NO₃⁻ | 71.5 |
| NH₄⁺ | 73.6 | SO₄²⁻ (per eq.) | 80.0 |
| ½Ca²⁺ | 59.5 | HCO₃⁻ | 44.5 |
| ½Mg²⁺ | 53.1 | CO₃²⁻ (per eq.) | 69.3 |
Table 1. Limiting equivalent ionic conductances at 25 °C (standard reference values, e.g. from Robinson & Stokes/CRC compilations). Divalent ions listed per equivalent so the Kohlrausch sum uses directly.
Goal: compute for a 4.50 mM Na₂CO₃ / 1.40 mM NaHCO₃ eluent, the kind commonly used for anion-exchange separations, before any suppression — to see just how large the problem is.
Step 1 — enumerate all ionic species and their concentrations. Full dissociation of the salts (both are strong electrolytes as salts, even though the acid forms are weak) gives:
(We neglect the small extent of CO₃²⁻ hydrolysis to HCO₃⁻/OH⁻ here; at these concentrations and pH ≈ 10.3, HCO₃⁻ dominates the buffer but the analytical concentrations already fix total carbonate species, and treating them as fully dissociated strong-electrolyte contributions is the standard first approximation for background κ.)
Step 2 — apply Kohlrausch's law term by term. For Na⁺ ():
To keep units straight, work in mol/cm³ and convert at the end. Using in mol/L and the standard shortcut (this shortcut works because mM × (S·cm²/mol) numerically gives µS/cm directly — mol/L × 10⁻³ × S·cm²/mol × 10⁶ µS/S = µS/cm when C is entered in mM):
(Here for CO₃²⁻ uses in mM and , consistent with the "per equivalent" already including the charge factor — so we multiply the molar concentration by 2 explicitly since the table value 69.3 is quoted per equivalent, i.e., already normalized; multiplying by converts molar to equivalent concentration.)
Step 3 — sum the contributions.
Step 4 — sanity-check magnitude. Real carbonate/bicarbonate eluents at these concentrations typically show measured background conductivities of several hundred to ~1500 µS/cm depending on exact composition and temperature, consistent with this estimate; the calculation correctly shows the background is three orders of magnitude above a typical few-µS/cm analyte signal, motivating suppression.
Limiting case: if the eluent concentration → 0 (pure water eluent, as used with electrolytically generated hydroxide gradients and suppression designed for that mode), (the conductivity of pure water from its autoionization), confirming the formula degrades correctly to the trivial no-electrolyte limit.
Go deeper · Derivation B · 1 minQuantifying the suppression gain: background collapse and analyte amplification
Goal: show algebraically why suppression simultaneously shrinks background and grows signal, using the same 4.50 mM Na₂CO₃/1.40 mM NaHCO₃ eluent now carrying 2.00 mg/L Cl⁻ as analyte (as NaCl), and quantify the improvement in signal-to-background ratio (S/B).
Step 1 — post-suppression background: carbonate becomes carbonic acid. After exchanging every Na⁺ for H⁺, the anions CO₃²⁻ and HCO₃⁻ combine with the liberated H⁺ to form H₂CO₃ (aq), a weak acid with . Total carbonate-system concentration is:
Treating this as an undissociated weak acid at first pass, the fraction ionized is estimated from
so .
Step 2 — background κ after suppression from carbonic acid alone.
This is already a ~60-fold drop from 1207 µS/cm, and real membrane suppressors do even better because dissolved CO₂ partly degasses or is swept out, and modern suppressors achieve residual backgrounds of 0.5–2 µS/cm in practice; our equilibrium-only estimate captures the chemical (not physical) part of the improvement and is a conservative upper bound.
Step 3 — analyte signal before suppression (as NaCl). Convert 2.00 mg/L Cl⁻ (M = 35.45 g/mol) to concentration:
As NaCl (Na⁺ and Cl⁻ both present, riding on top of the eluent's own Na⁺):
But this 7.13 µS/cm sits on a 1207 µS/cm background: S/B = 7.13/1207 = 0.0059 — under 0.6%, essentially undetectable against pump-noise-scaled baseline drift.
Step 4 — analyte signal after suppression (as HCl).
Now S/B = 24.0/19.9 = 1.21 — comparable to or exceeding the (much smaller) background, a roughly 200-fold improvement in S/B compared to Step 3.
Step 5 — decompose the total gain.
Limiting case: if the analyte were itself already a strong-acid/base pair needing no suppression gain (e.g., detecting H⁺ directly in a cation system), the amplification factor above would reduce toward 1, and the entire benefit of suppression would come from background reduction alone — confirming the two mechanisms (background drop, signal rise) are independent and additive in log-gain, not artifacts of each other.
Worked examples
Example 1 (Foundational): Converting a suppressed-mode chromatogram peak area to concentration via linear calibration. A lab runs external standards of NO₃⁻ at 0.500, 2.00, and 5.00 mg/L under suppressed conductivity detection, obtaining peak areas (µS·min) of 0.842, 3.35, and 8.41 respectively. A sample gives a peak area of 2.10 µS·min. Find the NO₃⁻ concentration, with an honest uncertainty estimate. Step 1 — linear regression. Fit peak area (µS·min) vs. concentration (mg/L):
Deviations: : ; : .
The intercept is essentially zero (as expected for a well-suppressed baseline), giving confidence the calibration is not distorted by a residual background offset. Step 2 — invert for the sample.
Step 3 — report with appropriate significant figures. Given three calibration points (only 1 degree of freedom for residuals), a full confidence interval requires the standard IC/HPLC calibration-uncertainty propagation, but to the precision supportable by three points, report:
Limiting-case check: if the unknown's peak area exactly equaled a standard's (say 3.35 µS·min), the back-calculated concentration should recover 2.00 mg/L exactly: mg/L ✓, confirming the regression is self-consistent.
Example 2 (Applied): Estimating suppressor capacity lifetime and diagnosing a "breakthrough" peak. A packed-bed (or membrane) cation suppressor used for anion IC has an effective exchange capacity equivalent to continuously neutralizing 50.0 µequiv/min of incoming Na⁺ at the eluent flow used. The lab runs a Na₂CO₃/NaHCO₃ eluent identical to the one above (10.4 mM total Na⁺) at a flow rate of 1.00 mL/min for an unusually long unattended overnight sequence. Estimate how long until the suppressor's capacity is (in principle) exceeded, assuming no electrolytic regeneration (i.e., treating it as a finite, non-regenerated bed, as older "hand-regenerated" fiber suppressors required). Step 1 — compute the molar flux of Na⁺ arriving at the suppressor.
Since Na⁺ is monovalent, this equals 10.4 µequiv/min. Step 2 — compare to stated suppressor throughput capacity. The suppressor is rated at 50.0 µequiv/min of sustained exchange capacity (this is a steady-state regeneration rate for an electrolytic suppressor, or a total finite capacity divided by an assumed usable lifetime for a non-regenerated one — here we treat the number as a finite total capacity in µequiv for the non-regenerated case, e.g., total capacity budget , a realistic order of magnitude for a small fiber suppressor). Step 3 — estimate time to exhaustion.
Step 4 — interpret the diagnostic symptom. As the suppressor approaches exhaustion, an increasing fraction of Na⁺ is no longer exchanged for H⁺; the background baseline conductivity rises progressively (since unconverted Na₂CO₃/NaHCO₃ contributes far more than H₂CO₃), and analyte peaks can show reduced amplification and peak-shape distortion (fronting/tailing) as the local suppressor chemistry becomes spatially non-uniform along the bed — an easily recognized symptom in an overnight batch sequence where late-run samples show systematically low recoveries and elevated baseline noise.
Limiting-case check: a modern electrolytically self-regenerating suppressor has, in effect, because H⁺ (or OH⁻) is continuously generated from water by the applied current rather than drawn from a finite chemical reservoir; setting in Step 3 correctly gives , matching the real-world advantage of continuous electrolytic regeneration over finite packed-bed exchange capacity.
Example 3 (Exam/competition level): Resolving co-elution using selectivity and computing minimum resolvable concentration ratio from suppressed-mode noise. A suppressed anion-exchange separation resolves Cl⁻ and NO₂⁻ with resolution under standard conditions (peaks nearly Gaussian, comparable widths). The baseline noise (peak-to-peak) in suppressed conductivity mode is , and the detector's linear response constant (from calibration, analogous to Example 1's slope but per unit height rather than area) is for NO₂⁻ peak height. A regulatory method requires a limit of detection (LOD) defined as signal-to-noise = 3, and a limit of quantitation (LOQ) at S/N = 10. Additionally, determine the maximum concentration ratio that still allows NO₂⁻ to be quantified without its peak being obscured, given that at the peak-valley overlap contributes an interference equivalent to 4.0% of the larger peak's height superimposed on the smaller peak. Step 1 — LOD and LOQ concentrations from S/N criteria.
Step 2 — express the resolution-driven interference constraint. If Cl⁻ is present at concentration giving peak height (with , scaled from relative to the NO₂⁻ constant — here we take it as given/measured directly at ), the interference riding under the NO₂⁻ peak is:
For NO₂⁻ to be quantifiable at its own LOQ, we require the true NO₂⁻ signal to exceed the interference by at least the same S/N=10 margin used for LOQ — i.e., the interference itself must not exceed the noise-equivalent contribution already budgeted, so a conservative working criterion is that interference stays below the LOD signal (0.036 µS), ensuring it does not itself introduce a spurious quantifiable "peak" mistaken for signal:
Step 3 — express as a maximum tolerable ratio at the NO₂⁻ LOQ. At :
So Cl⁻ can outnumber NO₂⁻ by up to roughly 6.6-fold (by mass concentration) at the NO₂⁻ LOQ before peak overlap at compromises the quantitation; above this ratio, either better resolution (longer column, shallower gradient) or a different selectivity (different eluent concentration/column chemistry) is needed.
Limiting-case check: if resolution improved to baseline separation (, interference fraction → 0), the constraint in Step 2 relaxes to in this idealized model — i.e., no ratio limit from peak overlap remains, matching the physical expectation that fully resolved peaks impose no such mutual constraint (real chromatograms would still have some finite tailing contribution, but the model's leading term correctly vanishes).
Limiting cases, scope & common traps
| Situation | Behavior | Why |
|---|---|---|
| Eluent concentration → 0 | Background κ → κ(pure water) ≈ 0.055 µS/cm | Kohlrausch sum has no electrolyte terms left; only autoionization remains |
| Fully exhausted suppressor | Background κ rises toward unsuppressed eluent value | Suppressor can no longer exchange counter-ions; behaves as if absent |
| Weak-acid analyte (e.g., F⁻ at very low pH) | Suppression can convert analyte itself into a weak acid, reducing sensitivity | HF has ; suppressed fluoride signal is smaller than fully dissociated strong-acid analytes at equal concentration |
| Divalent analyte (SO₄²⁻, CO₃²⁻) | Larger conductivity contribution per mole than monovalent at same molar concentration | Kohlrausch term scales with ; charge, not just mobility, sets the equivalent conductance weight |
| Electrolytic (self-regenerating) suppressor at very high current | Excess water electrolysis can generate baseline instability / bubbles | Gas evolution (H₂/O₂) at electrodes must be managed by membrane design and flow geometry |
The chemist behind it
Suppressed ion chromatography traces to work at Dow Chemical in the mid-1970s by Hamish Small, Timothy Stevens, and William Bauman, who first demonstrated that a second ion-exchange column (or later, a membrane device) placed after the separator column could chemically neutralize eluent conductivity while amplifying analyte signal — turning conductivity detection, previously considered too nonspecific for trace ion analysis, into a viable quantitative tool. Their 1975 paper in Analytical Chemistry is generally credited as the founding publication of modern ion chromatography, and the technique was commercialized rapidly (Dionex Corporation, founded soon after specifically to exploit this discovery), transforming anion analysis from slow, wet-chemical, or specific-ion-electrode methods into a fast, multiplexed instrumental technique.
Origins · 1 minHamish Small and how this was discovered
Small was an analytical chemist working on ion-exchange separations at Dow when he recognized that a "stripper" or suppressor column placed in series after a separator column could remove the ionic strength of an eluent buffer while leaving the separated analyte ions in a conductivity-enhanced form. Working with Stevens and Bauman, he built the first packed-bed suppressor systems and demonstrated quantitative anion determinations that had previously required laborious titrimetric or gravimetric methods. The suppression concept — separating what elutes the ions from what you detect them as — remains the conceptual core of essentially every commercial IC system sold today.
Why can't we just use UV detection for inorganic anions like chloride and sulfate?
Most simple inorganic anions relevant to environmental and industrial IC — chloride, sulfate, fluoride, phosphate — have no meaningful UV chromophore; they absorb only deep in the vacuum-UV region inaccessible to standard HPLC detectors. Nitrate and nitrite are exceptions, showing usable UV absorbance near 210 nm, but this is not general. Conductivity detection, by contrast, responds to virtually any ionic species regardless of its electronic structure, since it measures charge transport rather than photon absorption. The tradeoff is that conductivity detection has no inherent selectivity — every ion contributes — which is exactly the problem suppression solves by removing the eluent's dominant contribution while leaving analyte ions detectable against a low, stable background.
Does suppression change the retention times or resolution of the separation?
No — suppression happens entirely after the separator column, in a post-column device with minimal additional dead volume in well-designed systems. The chromatographic separation (retention order, resolution, peak shape from the column itself) is set upstream, on the ion-exchange stationary phase, before suppression occurs. A poorly designed suppressor with excessive internal volume can, however, add band-broadening (extra-column dispersion) that degrades observed resolution even though the underlying separation was fine — this is why modern membrane suppressors are engineered with very small internal volumes (microliters), specifically to avoid degrading resolution achieved on the column.
Why is methanesulfonic acid (MSA) commonly used as the eluent for cation IC instead of, say, HCl or HNO₃?
MSA is a strong organic acid that, after suppression (using an anion-exchange suppressor in OH⁻ form), is converted into water plus methanesulfonate anion paired with whatever counter-ion was liberated — but critically, MSA itself, and the suppression byproduct, contribute very low background conductivity once neutralized, and MSA is compatible with gradient elution schemes used for separating mono- and divalent cations (Na⁺, K⁺, Mg²⁺, Ca²⁺) in one run. Strong mineral acids like HCl or HNO₃ can also be used but MSA's physical properties (lower volatility, less corrosive to system components, cleaner suppression chemistry) make it the standard commercial choice.
What sets the ultimate detection limit in suppressed conductivity IC if suppression removes almost all the background?
Once chemical background is suppressed to its practical floor (often 0.5–2 µS/cm for a well-maintained system), the limiting noise sources become instrumental: pump-flow pulsation (pressure/flow ripple modulating the tiny residual background), temperature fluctuations in the detector cell (conductivity is strongly temperature-dependent, roughly 1.5–2%/°C for most electrolytes), and electronic noise in the conductivity measurement circuit itself. This is why modern suppressed-IC systems use actively temperature-controlled detector cells and low-pulsation pumps — the suppression chemistry has already done its job; the remaining limit is engineering, not chemistry.
Can suppression be used with gradient elution, where eluent concentration changes during the run?
Yes, and this is one of the major advantages of modern electrolytic suppressors combined with electrolytic eluent generation. Because the suppressor's H⁺ (or OH⁻) supply is generated on demand by applied current rather than drawn from a fixed chemical reservoir, the suppressor can track a changing eluent concentration throughout a gradient without running out of capacity partway through, provided the applied current is scaled appropriately to match the instantaneous eluent ion flux. This pairing (electrolytic eluent generation plus electrolytic suppression) is now standard in high-end IC systems and enables highly reproducible hydroxide gradients generated from pure water and a power supply alone.
Is suppression ever undesirable or skipped in ion chromatography?
Yes — "non-suppressed" or single-column IC is still used, particularly for cheaper instrumentation, for anions with very high values that would themselves become weak/undetectable acids upon suppression (making suppression counterproductive), or when a low-capacity ion-exchange column paired with a low-conductivity eluent (e.g., dilute phthalate or borate/gluconate buffers) already gives an acceptable background without a suppressor. Non-suppressed IC trades some sensitivity and background stability for simpler, cheaper instrumentation, and remains common in lower-throughput or cost-sensitive laboratory settings.
How does a suppressor avoid contaminating the analyte stream with the regenerant chemicals?
In membrane-based suppressors, an ion-exchange membrane (not a porous frit) separates the eluent stream from the regenerant compartment. The membrane is selectively permeable to one ion type (H⁺ in a cation-exchange membrane, OH⁻ in an anion-exchange membrane) via a fixed-charge polymer network, and Donnan exclusion prevents co-ions (and, practically, the bulk uncharged regenerant/product species) from crossing into the eluent stream. Only the exchanging ion crosses the membrane; the eluent stream never physically mixes with the regenerant liquid, which is why the regenerant compartment can even be pure deionized water undergoing electrolysis without contaminating the analyte flow path.