Medical Physics: X-Ray, CT, Ultrasound & Tracers

Within weeks of Wilhelm Röntgen's discovery of X-rays in 1895, he made a photograph that stunned the world: the bones of his wife's hand, her wedding ring floating dark around a finger, the flesh a faint grey ghost. For the first time a doctor could look inside a living body without a scalpel. Anna Röntgen is said to have gasped, "I have seen my death." She had also seen the birth of medical physics. Everything a modern hospital does to image you — the chest X-ray, the CT scanner that builds a cross-section of your skull, the ultrasound probe sliding over an expectant mother's abdomen, the gamma camera that watches a radioactive tracer light up a tumour — is applied physics you already half-know: the electromagnetic spectrum, the exponential decay law, wave reflection at a boundary, and radioactive decay, all pointed at the human body. This article gathers those threads into the four great windows into the living body: seeing by shadow (X-ray), by reconstructed slice (CT), by echo (ultrasound), and by emission (radionuclide tracers), with the electrocardiogram as a bonus window onto the heart. We keep to school algebra throughout. The law that does the heavy lifting for X-rays — the exponential attenuation I=I0e−μxI = I_0 e^{-\mu x} — is quoted, exactly as you already quote the radioactive-decay law N=N0e−λtN = N_0 e^{-\lambda t}; the half-value thickness follows by the same arithmetic. Ultrasound needs only wave reflection and the definition of acoustic impedance. The reward is understanding, from first principles, how physics learned to see through skin.

Key takeaways
  • Medical imaging is four kinds of "seeing": transmission (X-ray), reconstruction (CT), reflection (ultrasound), and emission (radionuclide tracers) — each exploits physics you already know.
  • X-rays make a shadow. Bone absorbs far more than soft tissue, so the transmitted beam carries a silhouette; the intensity falls as I=I0e−μxI = I_0 e^{-\mu x}, the same exponential form as radioactive decay.
  • The half-value thickness HVT=0.693/μ\text{HVT} = 0.693/\mu is the layer that halves the beam. After nn such layers the intensity is I0/2nI_0/2^n: three HVTs leave 12.5%12.5\%.
  • CT turns many X-ray shadows into a true cross-sectional slice, each pixel graded in Hounsfield units: water 00, air −1000-1000, dense bone about +1000+1000.
  • Ultrasound is non-ionising and works by echo. A boundary reflects sound in proportion to the mismatch in acoustic impedance Z=ρcZ = \rho c; the tissue–air mismatch reflects about 99.9%99.9\%, hence coupling gel.
  • Radionuclide imaging shows function, not just anatomy — a tracer such as technetium-99m (half-life 66 h, 140 keV140\ \text{keV} gamma) is injected and its emission mapped by a gamma camera or PET.
  • Key idea: every medical image is the electromagnetic spectrum, the exponential attenuation law, wave reflection at an impedance boundary, or radioactive decay — turned on the human body, read with nothing harder than algebra.
The core insight

Medical imaging sees inside the body four ways — by the shadow a differentially-absorbed X-ray beam casts (I=I0e−μxI = I_0 e^{-\mu x}, halved every HVT=0.693/μ\text{HVT} = 0.693/\mu), by the slice CT reconstructs from many such shadows, by the echo ultrasound gets from mismatches in acoustic impedance Z=ρcZ = \rho c (tissue–air reflects ≈99.9%\approx 99.9\%, hence gel), and by the emission of an injected radioactive tracer — all standard-syllabus physics pointed at anatomy.

Where this lives today

Medical imaging is one of physics' quietest triumphs. A modern CT scanner rotates its X-ray tube around you in under a second and reconstructs a hundred slices; with iodine contrast it traces a blocked coronary artery without an incision. PET fused with CT overlays function on anatomy, lighting up an active tumour against the organs around it, and has reshaped how cancers are staged. Ultrasound has become the stethoscope of the twenty-first century — cheap, portable, harmless, now shrunk to a probe that plugs into a phone and images a heart at the bedside. The frontier now is computational. The physics of transmission, reflection, and emission is settled; what advances fast is the reconstruction — machine-learning algorithms that build a clean image from a lower dose, de-noise a fast ultrasound sweep, or spot a lesion a tired eye might miss. Hounsfield's revolution was to realise that an image need not be captured directly — it can be computed from indirect measurements, an idea, born in a lab funded by pop records, that still drives the field.

The physical picture — four windows into the living body

The body is not transparent, but it is not opaque either: different tissues let different kinds of radiation through by different amounts, and that difference is the image. A picture forms wherever one structure absorbs, reflects, or emits differently from its neighbour — contrast, not brightness, is what a radiologist reads. Imaging comes down to finding a signal that bone treats differently from muscle, or a tumour differently from healthy tissue. There are exactly four ways to get such a signal, and every scanner is one of them. Send radiation through the body and record what survives — that is transmission, what a plain X-ray film does. Send it through from many angles and let a computer reconstruct a true slice — that is computed tomography. Send a pulse in and listen for what bounces back — that is ultrasound, reflection imaging. Or put a source inside the body, as an injected tracer, and record what comes out — that is radionuclide (nuclear-medicine) imaging. Transmission, reconstruction, reflection, emission: four windows, four pieces of physics.

Figure
The four windows of medical imaging: an X-ray beam transmitted through the body to make a shadow, a CT scanner rotating to reconstruct a slice, an ultrasound probe sending a pulse and receiving an echo, and a gamma camera detecting radiation emitted by an injected tracer inside the body
Four ways to see inside the body 1 · Transmission X-ray shadow S bone absorbs film 2 · Reconstruction CT slice S source rotates many angles → slice 3 · Reflection ultrasound echo probe pulse in echo back impedance boundary time of echo → depth non-ionising 4 · Emission radionuclide tracer γ camera source is inside shows function

Two of these windows use ionising radiation — X-ray, CT, and the tracer's gamma rays carry enough energy per photon to knock electrons out of atoms, so they deliver a dose and are used with care. Two do not: ultrasound is a mechanical (sound) wave and the electrocardiogram merely listens to the heart's own voltages. This ionising-versus-not divide governs which modality a doctor reaches for — you would never CT a healthy fetus but would scan it with ultrasound, precisely because ultrasound deposits no ionising dose. It is the single most examined concept in the whole topic.

Seeing by shadow: X-rays and the attenuation law

An X-ray image is a shadow. Behind the patient sits a detector (film, or a flat digital panel); in front, an X-ray tube fires a broad beam of high-energy photons — diagnostic energies of about 3030 to 120 keV120\ \text{keV}, produced as in the previous article on X-ray production. As the beam crosses the body, every tissue removes some photons (by the photoelectric effect and Compton scattering), and dense, high–atomic-number material such as bone removes far more than soft tissue of the same thickness. So the beam reaching the detector is strong where it passed through flesh and weak where it passed through bone — and the developed image is the negative of that: bone prints white, air-filled lung dark, soft tissue a mid grey. The picture maps how much of the beam each straight-line path let through. How fast does a beam die away inside a uniform absorber? For a narrow monoenergetic beam the answer is the exponential attenuation law, which we quote — it has the same standing, and the same shape, as the radioactive-decay law N=N0e−λtN = N_0 e^{-\lambda t} you already use in nuclear physics:

I=I0 e−μx.(1)I = I_0\, e^{-\mu x}. \qquad (1)

where: I0I_0 is the beam intensity entering the absorber, II the intensity after a depth xx, and μ\mu the linear attenuation coefficient (units m−1\text{m}^{-1} or cm−1\text{cm}^{-1}), a property of both the material and the photon energy. The physical reading: each thin layer removes the same fraction of whatever beam reaches it, so the intensity decays geometrically with depth, never quite reaching zero. A large μ\mu (bone, or a contrast agent like iodine or barium) means a beam that dies quickly and casts a strong shadow; a small μ\mu (soft tissue, lung) leaves little contrast. This energy-dependence is deliberate clinical currency: physicians pick a beam energy that maximises the difference in μ\mu between the structures they want to tell apart. Because equation (1) is exponential, the natural way to describe an absorber is not "how much stops the beam" but "how much halves it" — the half-value thickness (HVT, or half-value layer). Setting I=I0/2I = I_0/2 in equation (1) gives a relation identical in form to the radioactive half-life t1/2=0.693/λt_{1/2} = 0.693/\lambda:

HVT=ln⁡2μ=0.693μ,I=I02 n after n half-value thicknesses.(2)\text{HVT} = \frac{\ln 2}{\mu} = \frac{0.693}{\mu}, \qquad I = \frac{I_0}{2^{\,n}}\ \text{after } n\ \text{half-value thicknesses.} \qquad (2)

where: ln⁡2=0.693\ln 2 = 0.693, μ\mu is the same linear attenuation coefficient, and n=x/HVTn = x/\text{HVT} is how many half-value layers deep you have gone. The physical reading is the whole reason HVT is used: it converts an awkward exponential into simple halving. One HVT leaves 50%50\%, two leaves 25%25\%, three leaves 12.5%12.5\% — you never need a calculator, only powers of one-half. Radiographers quote shielding in HVTs for exactly this reason.

Figure
A monoenergetic X-ray beam of intensity I nought entering tissue and bone: the transmitted intensity falls exponentially with depth, dropping to one half after one half-value thickness, one quarter after two, and one eighth after three, while the denser bone with a larger attenuation coefficient drops the beam far faster and casts the strong white shadow
The beam dies exponentially — halved every half-value thickness I depth x I₀ I₀/2 I₀/4 I₀/8 soft tissue · small μ bone · large μ 1 HVT 2 HVT 3 HVT 50% 25% 12.5% Each half-value thickness removes half of whatever beam remains · bone empties the beam in a fraction of the distance → white shadow

From shadow to slice: computed tomography

A plain X-ray has one great weakness: it collapses the whole depth of the body onto one plane, so a small tumour hidden behind a rib can be invisible. Godfrey Hounsfield's insight, built at EMI in 1971, was that if you record the transmitted beam from many directions around the patient and feed the set to a computer, the machine can solve for the attenuation coefficient at every point of a thin cross-sectional slice — a genuine map of the interior, one plane at a time. That is computed tomography, from the Greek tomos, a slice. The mathematics of reconstructing an object from all its projections had been worked out by Johann Radon in 1917; Hounsfield and, independently, Allan Cormack turned it into a machine. CT does not report raw μ\mu; it reports a calibrated number, the Hounsfield unit (HU), comparing each voxel's attenuation to water's:

HU=1000×μ−μwaterμwater.(3)\text{HU} = 1000\times\frac{\mu - \mu_{\text{water}}}{\mu_{\text{water}}}. \qquad (3)

where: μ\mu is the local attenuation coefficient and μwater\mu_{\text{water}} that of water at the same beam energy. The physical reading is a clinician-friendly scale pinned to two anchors: water sits at exactly 0 HU0\ \text{HU} by construction, and air (with essentially no attenuation, μ≈0\mu \approx 0) at −1000 HU-1000\ \text{HU}. Dense cortical bone, whose μ\mu is roughly twice water's, lands near +1000 HU+1000\ \text{HU}; fat around −100 HU-100\ \text{HU}; clotted blood around +60 HU+60\ \text{HU}. A single CT slice measures μ\mu everywhere; the Hounsfield unit is just how that measurement is packaged.

Seeing by echo: ultrasound and acoustic impedance

Switch off the radiation entirely. Ultrasound sends a pulse of sound — mechanical vibration far above hearing, typically 11 to 15 MHz15\ \text{MHz} — into the body from a hand-held transducer, then falls silent and listens. Wherever the pulse crosses a boundary between tissues, part reflects straight back as an echo; the transducer times each echo and, knowing the speed of sound in soft tissue (v≈1540 m/sv \approx 1540\ \text{m/s}), converts the delay into a depth. Sweep the beam across a plane and the machine paints a real-time cross-section. Because sound is not ionising, ultrasound carries no radiation dose — the reason it, and not X-rays, is used to watch a developing baby. What decides how strong an echo a boundary gives? Not density alone and not stiffness alone, but a combination of the two called the acoustic impedance:

Z=ρc.(4)Z = \rho c. \qquad (4)

where: ρ\rho is the tissue's density and cc the speed of sound in it (units of ZZ: kg m−2s−1\text{kg}\,\text{m}^{-2}\text{s}^{-1}, often written as the rayl). The physical reading: acoustic impedance measures how much a medium "resists" being set into sound-vibration. Soft tissue has Z≈1.63×106Z \approx 1.63\times10^{6} (density about 1060 kg/m31060\ \text{kg/m}^3, sound speed 1540 m/s1540\ \text{m/s}); air is far lower, about 430430; bone far higher, about 7.8×1067.8\times10^{6}. An echo appears only where ZZ changes, and its strength depends on the size of the jump. The fraction of intensity reflected at a boundary between media of impedance Z1Z_1 and Z2Z_2 is:

R=(Z2−Z1Z2+Z1)2.(5)R = \left(\frac{Z_2 - Z_1}{Z_2 + Z_1}\right)^2. \qquad (5)

where: Z1Z_1 and Z2Z_2 are the impedances on the two sides of the boundary, and RR is the reflected fraction of the incident intensity (the rest transmits onward to make deeper echoes). The physical reading holds the two facts every exam wants. First, a matched boundary (Z2=Z1Z_2 = Z_1) reflects nothing — it is acoustically invisible — while boundaries between soft tissues, differing in impedance by only a percent or two, give faint echoes of about 1%1\% that are exactly what a scan is built from. Second, a mismatched boundary reflects almost everything: at a tissue–air interface the impedances differ by a factor of nearly four thousand, so R≈99.9%R \approx 99.9\% — virtually the whole pulse bounces off the first sliver of air. That is why the sonographer smears coupling gel between probe and skin: it expels the thin air layer that would otherwise mirror the beam straight back. The same near-total reflection at tissue–bone and tissue–gas boundaries is why ultrasound cannot see through lungs or behind bone.

Figure
An ultrasound transducer coupled to skin through gel sends a pulse into the body; at each impedance boundary a fraction R equals the square of Z two minus Z one over Z two plus Z one reflects back, giving a faint one percent echo at a soft-tissue boundary but a near total ninety-nine point nine percent reflection at any tissue to air boundary, which is why gel is needed
Ultrasound echoes: a boundary reflects in proportion to its impedance mismatch probe gel (matches probe to skin) soft tissue Z ≈ 1.63 × 10⁶ other soft tissue Z ≈ 1.70 × 10⁶ bone Z ≈ 7.8 × 10⁶ pulse in R ≈ 1% weak echo R ≈ 43% strong echo, shadows behind No gel: a trapped air film tissue Z ≈ 1.63 × 10⁶ air Z ≈ 430 R ≈ 99.9% — almost all reflects nothing reaches the tissue below → smear gel to expel the air

One more number matters: resolution. Like any wave, sound cannot resolve detail finer than about one wavelength, λ=v/f\lambda = v/f. At 2 MHz2\ \text{MHz} in tissue λ=1540/(2×106)=0.77 mm\lambda = 1540/(2\times10^6) = 0.77\ \text{mm}; at 10 MHz10\ \text{MHz} it shrinks to 0.15 mm0.15\ \text{mm}, sharpening the image. But higher frequencies are absorbed faster and penetrate less deeply, so a sonographer trades resolution against reach: a high-frequency probe for a shallow thyroid, a lower-frequency one for a deep fetus or liver.

Seeing by emission: radionuclide imaging and the ECG

The three modalities so far show anatomy — what is where. Nuclear medicine shows function — what is working. A radionuclide tracer is chemically tailored so the body takes it to a particular organ or process — a bone-seeking compound to the skeleton, a sugar analogue to hungry tumour cells — and, being radioactive, it broadcasts its own location. A gamma camera maps where the emitted gamma rays come from; positron-emission tomography (PET) does the same with positron emitters, detecting the two back-to-back annihilation photons. The image is not of density but of activity: a "hot spot" is a region taking up the tracer avidly. The workhorse tracer is technetium-99m, and its choice is a lesson in radiation physics. Its half-life is about 66 hours — long enough to inject, circulate, and scan, short enough to decay away within a day and spare the dose. It emits a single 140 keV140\ \text{keV} gamma — energetic enough to escape the body yet low enough to capture efficiently — and, being an isomeric ("m", metastable) transition, essentially no alpha or beta, which would add dose without adding image. It is the same α\alpha/β\beta/γ\gamma physics you met for radioactive decay, now chosen deliberately: a pure, medium-energy gamma emitter with a short half-life is the ideal messenger. The electrocardiogram (ECG) is the odd one out — no image, no radiation. Every heartbeat is triggered by a wave of electrical depolarisation across the heart muscle, setting up tiny voltages, of order a millivolt, on the skin. Electrodes record the familiar trace: the small P wave (atria), the sharp QRS complex (ventricles firing), the rounded T wave (ventricles resetting). From the peak spacing a cardiologist reads the heart rate — a healthy 7575 beats per minute is one beat every 0.8 s0.8\ \text{s} — and from their shape, the health of the muscle and its conduction. It sits in medical-physics syllabuses because it is pure applied electricity: potentials, electrodes, and filtering out the 50 Hz50\ \text{Hz} mains hum.

Derivation AThe half-value thickness: how many centimetres halve the beam

The goal is to pin down, from the quoted attenuation law alone, both the size of one half-value thickness for a given absorber and the fraction of beam left after any number of them — turning the exponential I=I0e−μxI = I_0 e^{-\mu x} into the pocket rule "I0/2nI_0/2^n". We use only logarithms and powers of one-half; no calculus, because equation (1) is quoted just as the radioactive-decay law is.

Step 1. Define the half-value thickness as the depth x=HVTx = \text{HVT} where I=I0/2I = I_0/2. Substitute into I=I0e−μxI = I_0 e^{-\mu x}: 12=e−μ HVT\tfrac{1}{2} = e^{-\mu\,\text{HVT}}. Take natural logs: −ln⁡2=−μ HVT-\ln 2 = -\mu\,\text{HVT}, so HVT=ln⁡2/μ=0.693/μ\text{HVT} = \ln 2 / \mu = 0.693/\mu. This is identical in form to the radioactive half-life t1/2=0.693/λt_{1/2} = 0.693/\lambda — the same algebra, with depth playing the role of time and μ\mu that of the decay constant.

Step 2. Put a number to it. For a material with μ=0.693 cm−1\mu = 0.693\ \text{cm}^{-1} (a convenient teaching value), HVT=0.693/0.693=1.00 cm\text{HVT} = 0.693/0.693 = 1.00\ \text{cm}: one centimetre halves the beam. Check against the exponential directly — e−0.693×1=0.500e^{-0.693\times 1} = 0.500.

Step 3. Now stack the layers. After nn half-value thicknesses the beam has been halved nn times, so I/I0=(1/2)n=1/2nI/I_0 = (1/2)^n = 1/2^n. For n=1,2,3n = 1, 2, 3 this gives 0.500, 0.250, 0.1250.500,\ 0.250,\ 0.125 — that is 50%50\%, 25%25\%, 12.5%12.5\%. Three HVTs leave one-eighth.

Step 4. A non-integer thickness works the same way. Through x=2.5 cmx = 2.5\ \text{cm}, n=x/HVT=2.5n = x/\text{HVT} = 2.5, so I/I0=2−2.5=0.177I/I_0 = 2^{-2.5} = 0.177 — about 17.7%17.7\% transmitted. The exponential agrees exactly: e−0.693×2.5=0.177e^{-0.693\times 2.5} = 0.177. The two roads are one and the same law. Verified: HVT =0.693/0.693=1.00 cm= 0.693/0.693 = 1.00\ \text{cm}; 1/2n=0.500, 0.250, 0.1251/2^n = 0.500,\,0.250,\,0.125 for n=1,2,3n=1,2,3; 2−2.5=e−0.693×2.5=0.1772^{-2.5} = e^{-0.693\times2.5} = 0.177 — recomputed in verify_hs2-medical-physics-imaging.py.

Go deeper · Derivation B · 1 minWhy the probe needs gel: the tissue–air reflection
Derivation BWhy the probe needs gel: the tissue–air reflection

The goal is to show, with the quoted reflection formula and the definition of acoustic impedance, that a tissue–air boundary reflects about 99.9%99.9\% of an ultrasound pulse — so an unimpeded air gap would blind the scan — while a soft-tissue boundary reflects only about 1%1\%, which is the useful signal. Only arithmetic and one squaring are involved.

Step 1. Fix the impedances from Z=ρcZ = \rho c. Soft tissue: Ztissue=1060×1540=1.63×106Z_{\text{tissue}} = 1060 \times 1540 = 1.63\times10^{6}. Air, with ρ≈1.3 kg/m3\rho \approx 1.3\ \text{kg/m}^3 and c≈330 m/sc \approx 330\ \text{m/s}: Zair=1.3×330=430Z_{\text{air}} = 1.3\times 330 = 430. The two differ by a factor of nearly four thousand.

Step 2. Feed them into R=((Z2−Z1)/(Z2+Z1))2R = \big((Z_2 - Z_1)/(Z_2 + Z_1)\big)^2 for tissue–air: R=((1.63×106−430)/(1.63×106+430))2=(0.99947)2=0.9989R = \big((1.63\times10^{6} - 430)/(1.63\times10^{6} + 430)\big)^2 = (0.99947)^2 = 0.9989 — 99.9%99.9\% reflected, only about one part in a thousand crossing an air gap. An air film between probe and skin mirrors the beam almost perfectly, so you would see nothing below it.

Step 3. Contrast a genuine soft-tissue boundary, fat (Z=1.38×106Z = 1.38\times10^{6}) against muscle (Z=1.70×106Z = 1.70\times10^{6}): R=((1.70−1.38)/(1.70+1.38))2=(0.104)2=0.0108R = \big((1.70 - 1.38)/(1.70 + 1.38)\big)^2 = (0.104)^2 = 0.0108 — about 1%1\%. Weak but perfectly detectable, and the other 99%99\% travels on to reveal deeper boundaries. This modest mismatch is what makes soft-tissue imaging possible.

Step 4. Read the design lesson. To let the pulse in, banish the air: the water-based gel has an impedance close to skin's, so the probe–gel–skin path is nearly matched and almost all the energy enters. Run the physics the other way and it explains ultrasound's blind spots — tissue–bone gives R=((7.8−1.63)/(7.8+1.63))2=0.43R = \big((7.8 - 1.63)/(7.8 + 1.63)\big)^2 = 0.43, so 43%43\% reflects and the shadow behind bone (and behind gas-filled lung and bowel) hides what lies beyond. Verified: Ztissue=1060×1540=1.63×106Z_{\text{tissue}} = 1060\times1540 = 1.63\times10^{6}, Zair=430Z_{\text{air}} = 430; tissue–air R=0.9989R = 0.9989 (99.9%), fat–muscle R=0.011R = 0.011 (about 1%), tissue–bone R=0.43R = 0.43 — recomputed in verify_hs2-medical-physics-imaging.py.

Limiting cases

caseresultmeaning
absorber vanishingly thin, μx→0\mu x \to 0I→I0I \to I_0beam passes untouched — soft tissue on a plain film gives little contrast
very large μ\mu (bone, iodine, barium)beam dies fast, strong white shadowwhy bones and contrast agents show; why lead shields
after nn half-value thicknessesI=I0/2nI = I_0/2^n50%, 25%, 12.5%50\%,\,25\%,\,12.5\% for n=1,2,3n=1,2,3
matched boundary Z2=Z1Z_2 = Z_1R=0R = 0acoustically invisible — no echo
tissue–air (or tissue–gas)R≈99.9%R \approx 99.9\%near-total reflection → needs gel; can't image past lung or bowel gas
tissue–boneR≈43%R \approx 43\%strong echo, acoustic shadow behind bone
higher ultrasound frequencyshorter λ\lambda, finer detail, less depththe resolution-versus-penetration trade-off
non-ionising modality (ultrasound, ECG)zero radiation dosethe choice for fetal and repeated imaging

Worked examples

Board level — half-value thickness and transmitted fraction (IB-HL and A-Level pattern)

An X-ray beam passes through an absorber with μ=0.693 cm−1\mu = 0.693\ \text{cm}^{-1}. (a) Find the half-value thickness. (b) What fraction is transmitted through 3.0 cm3.0\ \text{cm}? (c) As a percentage?

Step 1. From equation (2): HVT=0.693/μ=0.693/0.693=1.0 cm\text{HVT} = 0.693/\mu = 0.693/0.693 = 1.0\ \text{cm} — one centimetre halves the beam.

Step 2. Through 3.0 cm3.0\ \text{cm} the beam crosses n=x/HVT=3n = x/\text{HVT} = 3 half-value thicknesses, so I/I0=1/23=1/8=0.125I/I_0 = 1/2^3 = 1/8 = 0.125.

Step 3. That is 12.5%12.5\% transmitted, 87.5%87.5\% removed. Cross-check the raw exponential: e−0.693×3=e−2.079=0.125e^{-0.693\times3} = e^{-2.079} = 0.125. Answer: HVT =1.0 cm= 1.0\ \text{cm}; transmitted =0.125= 0.125, i.e. 12.5%12.5\%.

Board level — acoustic impedance and why we use gel (A-Level pattern)

Soft tissue has density 1060 kg/m31060\ \text{kg/m}^3 and sound speed 1540 m/s1540\ \text{m/s}. (a) Find its acoustic impedance. (b) A thin air layer (Zair≈430Z_{\text{air}} \approx 430) is trapped between probe and skin — what fraction of the pulse reflects, and what does that say about gel?

Step 1. From equation (4): Z=ρc=1060×1540=1.63×106 kg m−2s−1Z = \rho c = 1060 \times 1540 = 1.63\times10^{6}\ \text{kg}\,\text{m}^{-2}\text{s}^{-1}.

Step 2. From equation (5): R=((1.63×106−430)/(1.63×106+430))2=0.9989R = \big((1.63\times10^{6} - 430)/(1.63\times10^{6} + 430)\big)^2 = 0.9989.

Step 3. So about 99.9%99.9\% reflects off the air layer and only 0.1%0.1\% enters — the scan would be blank. Gel, whose impedance is close to tissue's, displaces the air and nearly matches probe to skin. Answer: Ztissue=1.63×106Z_{\text{tissue}} = 1.63\times10^{6}; R≈99.9%R \approx 99.9\%, so gel is essential.

Competitive level — echo depth from time of flight (A-Level and IB-HL pattern)

Speed of sound in soft tissue is 1540 m/s1540\ \text{m/s}. (a) An echo returns 100 μs100\ \mu\text{s} after the pulse — how deep is the boundary? (b) What is the wavelength and best resolution at 2 MHz2\ \text{MHz}? (c) Why switch to 10 MHz10\ \text{MHz}, and at what cost?

Step 1. The pulse goes down and the echo comes back up, so the sound covers 2d2d in time tt: d=v t/2=1540×(100×10−6)/2=0.077 m=7.7 cmd = v\,t/2 = 1540 \times (100\times10^{-6})/2 = 0.077\ \text{m} = 7.7\ \text{cm}.

Step 2. Resolution is set by wavelength: λ=v/f=1540/(2×106)=0.77 mm\lambda = v/f = 1540/(2\times10^{6}) = 0.77\ \text{mm}.

Step 3. At 10 MHz10\ \text{MHz}, λ=0.154 mm\lambda = 0.154\ \text{mm} — five times finer, but higher frequencies are absorbed faster and cannot reach deep, so the 10 MHz10\ \text{MHz} probe suits shallow targets only. Answer: d=7.7 cmd = 7.7\ \text{cm}; λ=0.77 mm\lambda = 0.77\ \text{mm} at 2 MHz2\ \text{MHz}, 0.15 mm0.15\ \text{mm} at 10 MHz10\ \text{MHz}. Timing note: under a minute. Halve the time-of-flight before multiplying (the round-trip factor of two is the most-forgotten step), and λ=v/f\lambda = v/f uses the same 1540 m/s1540\ \text{m/s} — this is sound, not light, so never reach for 3×1083\times10^8.

Competitive level — mixed attenuation and a tracer sanity check (A-Level pattern)

(a) An X-ray beam passes through 2.5 cm2.5\ \text{cm} of a material with μ=0.693 cm−1\mu = 0.693\ \text{cm}^{-1} — what fraction transmits? (b) A technetium-99m dose (half-life 6 h6\ \text{h}) is injected — what fraction remains after 24 h24\ \text{h}? (c) The tracer emits a 140 keV140\ \text{keV} gamma — estimate its wavelength.

Step 1. n=2.5/1.0=2.5n = 2.5/1.0 = 2.5 half-value thicknesses, so I/I0=2−2.5=0.177I/I_0 = 2^{-2.5} = 0.177, about 17.7%17.7\% (exponential check: e−0.693×2.5=0.177e^{-0.693\times2.5} = 0.177).

Step 2. Tracer decay is the same half-life rule: 24 h24\ \text{h} is 44 half-lives, so (1/2)4=1/16=0.0625(1/2)^4 = 1/16 = 0.0625, i.e. 6.25%6.25\% remains — nearly all gone within a day, why such a short half-life is chosen.

Step 3. From λ=hc/E\lambda = hc/E with hc=1240 eV nmhc = 1240\ \text{eV}\,\text{nm} and E=1.4×105 eVE = 1.4\times10^{5}\ \text{eV}: λ=1240/(1.4×105)=8.9×10−3 nm=8.9 pm\lambda = 1240/(1.4\times10^{5}) = 8.9\times10^{-3}\ \text{nm} = 8.9\ \text{pm} — a hard gamma, why it escapes the body. Answer: (a) 17.7%17.7\%; (b) 6.25%6.25\%; (c) λ≈8.9 pm\lambda \approx 8.9\ \text{pm}. Timing note: seconds each once you spot the structure — attenuation and radioactive decay are the SAME exponential, 2−x/HVT2^{-x/\text{HVT}} and 2−t/t1/22^{-t/t_{1/2}}, so count half-layers or half-lives and raise one-half to that power.

Exam traps

Calling a plain X-ray a "slice". A radiograph is a shadow — a projection of the whole depth onto one plane. Only CT resolves depth by reconstructing from many angles. This anatomy-versus-projection point is a standard IB-HL and A-Level discrimination question.

Adding half-value thicknesses instead of multiplying. Each HVT halves what remains, so nn layers leave 1/2n1/2^n, not 1−n/21 - n/2. Three HVTs leave 12.5%12.5\%, not "nothing" and not 25%25\%. The exponential is multiplicative — this is the same trap as radioactive half-lives.

Treating μ\mu as fixed. The linear attenuation coefficient depends on photon energy as well as material; changing the tube voltage changes the contrast. Do not quote a single μ\mu for "bone" as if it were a constant.

Confusing ionising and non-ionising modalities. X-ray, CT, and radionuclide/PET use ionising radiation and deliver a dose; ultrasound (sound) and ECG (voltages) do not. The reason ultrasound, not CT, is used in pregnancy is precisely this. A frequent slip is to call ultrasound or MRI "a type of X-ray" — it is not.

Forgetting the factor of two in pulse-echo. The echo makes a round trip, so depth is d=vt/2d = vt/2, not vtvt. And use the speed of sound (1540 m/s1540\ \text{m/s}), never the speed of light.

Board scope — read this carefully. Medical imaging is a genuine syllabus topic only on ICSE/ISC, IGCSE 0625, IB Physics HL, and A-Level (e.g. the AQA Medical Physics option). On IGCSE it is qualitative only — you are asked about the uses of ultrasound and radiation, not to compute an HVT or an impedance. The quantitative treatment here — I=I0e−μxI = I_0 e^{-\mu x}, HVT=0.693/μ\text{HVT} = 0.693/\mu, Z=ρcZ = \rho c, the reflection formula — is IB-HL and A-Level material. It is not in the CBSE, JEE, NEET, or AP Physics 2 core, so if you sit those, treat this as enrichment that deepens the X-ray, wave-reflection, and radioactive-decay physics you are examined on.

Origins · 1 minGodfrey Hounsfield and how this was discovered
GH
Godfrey Hounsfield
1919–2004 · EMI Central Research Laboratories, Hayes, England

Godfrey Newbold Hounsfield was an English electrical engineer who left school without the qualifications for university, learned electronics in the RAF during the Second World War, and spent his career at EMI — the company better known for recording the Beatles. It was partly Beatles revenue that funded his obsession: reconstructing the interior of an object from X-ray measurements taken all around it. In 1971 his prototype scanner produced the first clinical CT image of a patient's brain, revealing a lesion (a cyst) with a clarity no plain X-ray could match. He had turned Johann Radon's 1917 mathematics and a great deal of computing into a machine, and it transformed diagnosis overnight. Hounsfield shared the 1979 Nobel Prize in Physiology or Medicine with Allan Cormack, who had independently derived the reconstruction theory. The CT brightness scale — zero at water — bears his name: the Hounsfield unit.

History
1895
— Wilhelm Röntgen discovers X-rays; within weeks he records the first radiograph, the bones of his wife Anna's hand
1901
— Röntgen receives the very first Nobel Prize in Physics, for the discovery of X-rays
1917
— Johann Radon proves that an object can be reconstructed from the complete set of its projections — the mathematics computed tomography would later need
1958
— Ian Donald introduces pulse-echo ultrasound into obstetrics in Glasgow, opening the era of the prenatal scan
1971
— Godfrey Hounsfield performs the first clinical CT scan, of a brain, at Atkinson Morley's Hospital, London (the scanner having been built at EMI in Hayes)
1979
— Hounsfield and Allan Cormack share the Nobel Prize in Physiology or Medicine for developing computed tomography

Going further

Texts. IB Physics (Tsokos), the Imaging option · A-Level Physics (AQA), the Medical Physics topic (attenuation, HVT, acoustic impedance, gamma camera) · ISC Physics Class 12 for the ICSE treatment · Webb, Introduction to Biomedical Imaging, and Bushberg et al., The Essential Physics of Medical Imaging, for the full account.

Provenance. Constants from LANE_SPEC: sound in tissue 1540 m/s1540\ \text{m/s}; ln⁡2=0.693\ln 2 = 0.693; hc=1240 eV nmhc = 1240\ \text{eV}\,\text{nm}. Recompute list: HVT=0.693/μ=1.0 cm\text{HVT} = 0.693/\mu = 1.0\ \text{cm}; 1/2n=0.500, 0.250, 0.1251/2^n = 0.500,\,0.250,\,0.125 for n=1,2,3n=1,2,3 and 2−2.5=e−0.693×2.5=0.1772^{-2.5} = e^{-0.693\times2.5} = 0.177; Hounsfield units water 00, air −1000-1000, fat −100-100, bone +1000+1000 from HU=1000(μ−μw)/μw\text{HU} = 1000(\mu-\mu_w)/\mu_w; Z=ρcZ = \rho c giving tissue 1.63×1061.63\times10^{6}, air 430430; R=((Z2−Z1)/(Z2+Z1))2R = ((Z_2-Z_1)/(Z_2+Z_1))^2 giving tissue–air 0.99890.9989 (99.9%), fat–muscle 0.0110.011 (1%), tissue–bone 0.430.43; λ=v/f=0.77 mm\lambda = v/f = 0.77\ \text{mm} at 2 MHz2\ \text{MHz}, 0.15 mm0.15\ \text{mm} at 10 MHz10\ \text{MHz}; echo depth d=vt/2=7.7 cmd = vt/2 = 7.7\ \text{cm} for t=100 μst = 100\ \mu\text{s}; Tc-99m half-life 6 h6\ \text{h}, fraction after 24 h24\ \text{h} is 1/24=6.25%1/2^4 = 6.25\%, gamma 140 keV140\ \text{keV} with λ=hc/E=8.9 pm\lambda = hc/E = 8.9\ \text{pm}; 75 min−175\ \text{min}^{-1} gives a beat every 0.8 s0.8\ \text{s}; CT/X-ray dose ratio 7/0.02=3507/0.02 = 350.

Prerequisites. X-Rays: Production & Spectra — where the beam this article sends through the body is made. Leads to. Ultrasound deepens the wave-reflection physics of the echo, and Radioactivity: Alpha, Beta & Gamma underpins the tracer emission of nuclear medicine. For the reconstruction mathematics behind CT (the Radon transform and its inversion, which need calculus) and the detector physics of PET, the university seam on biomedical imaging goes deeper — enrichment only, never required for the boards above.

Why does an X-ray image look like a negative, with bone white?

Because bone removes more of the beam than soft tissue, so fewer X-rays reach the detector behind bone. On film, more exposure darkens the emulsion, so regions behind soft tissue turn dark and regions behind bone stay light — bone prints white. It is a map of what the beam did not get through: dense, high-atomic-number structures (bone, metal, iodine contrast) block the most and appear brightest, while air-filled lung lets almost everything through and appears darkest.

Is the exponential attenuation law the same thing as radioactive decay?

They are two instances of the same mathematics. Radioactive decay, N=N0e−λtN = N_0 e^{-\lambda t}, describes how the number of undecayed nuclei falls with time; attenuation, I=I0e−μxI = I_0 e^{-\mu x}, how a beam's intensity falls with depth. Both say each equal step removes the same fraction of what is left, so the quantity halves over a fixed interval — a half-life 0.693/λ0.693/\lambda in one case, a half-value thickness 0.693/μ0.693/\mu in the other. If you can do half-life problems you can already do attenuation problems; only the symbols change.

Why can't ultrasound image the brain or the lungs?

Because of impedance mismatches so large they mirror the pulse. The adult skull is bone, and a tissue–bone boundary reflects about 43%43\% of the ultrasound and scatters much of the rest, so little usable signal reaches the brain. The lungs and bowel are full of gas, and a tissue–air boundary reflects about 99.9%99.9\% — essentially all of it — leaving an impenetrable acoustic shadow. Ultrasound therefore excels at soft-tissue targets reached without crossing bone or gas: the fetus in its fluid, the heart through the gaps between ribs, the liver, blood vessels. For the brain or lungs, doctors turn to CT or MRI.

Is a CT scan dangerous? How much radiation is it?

A CT uses a meaningfully larger dose than a single X-ray because it fires the beam from all around you. A chest X-ray is roughly 0.02 mSv0.02\ \text{mSv}; a chest CT is of order 7 mSv7\ \text{mSv} — very roughly 350350 times more, and comparable to a few years of natural background (about 2.4 mSv2.4\ \text{mSv} per year). That is why CT is not used casually, and never in place of an ultrasound when one will do — especially in children and pregnancy. But when a CT answers a serious question, the tiny long-term risk is overwhelmingly outweighed by the benefit of an accurate diagnosis.

Why choose technetium-99m and not some longer-lived tracer?

Everything about it is tuned for imaging. Its six-hour half-life is long enough to prepare, inject, and scan, yet short enough that it has almost entirely decayed within a day, so the dose is small and brief. Its 140 keV140\ \text{keV} gamma is high enough to escape the body but low enough to detect efficiently. And because it decays by an isomeric transition it emits essentially a pure gamma, with negligible alpha or beta that would add dose without adding image. A long-lived or particle-emitting nuclide would irradiate the patient for far longer to no diagnostic gain.

What does an ECG actually measure, and why is it in a physics syllabus?

It measures voltages — the potential differences, around a millivolt, that appear on the skin as a wave of electrical depolarisation sweeps across the heart with every beat. Electrodes draw the trace: the P wave (atria), the QRS complex (ventricles firing), the T wave (ventricles resetting). It sits in medical-physics courses because it is pure applied electricity — potential difference, electrodes, amplification, and filtering out the 50 Hz50\ \text{Hz} mains hum that would otherwise swamp a millivolt signal. From the peak spacing you get the heart rate (a beat every 0.8 s0.8\ \text{s} at 7575 per minute); from their shape, the health of the muscle and wiring.

Before this