PMMC instruments, shunts, and multipliers

A permanent-magnet moving-coil (PMMC) movement is the tiny, delicate mechanism hiding inside almost every analogue DC ammeter and voltmeter you have ever seen — a light coil of wire, suspended in a magnet's field, that swings a pointer in exact proportion to the current through it. Left alone, that coil can safely carry only a few milliamperes before it burns out, which seems to make it useless for measuring the tens or hundreds of amperes flowing in a real circuit, or the hundreds of volts across a real supply. The fix is two humble resistors. A shunt, wired in parallel with the coil, diverts almost all of a large current around the coil so only a safe trickle passes through it — turning the movement into an ammeter of any range you like. A multiplier, wired in series, forces the coil to see only a safe fraction of a large voltage — turning it into a voltmeter. Both designs fall straight out of Ohm's law and Kirchhoff's laws applied to the coil-plus-resistor combination, and this article works through the torque mechanism that makes the coil deflect proportionally to current, then derives and applies both design formulas.

The core insight

A PMMC coil deflects linearly with DC current because Td=NBAIT_d = NBAI balances a spring torque Tc=KθT_c = K\theta; a low parallel shunt Rsh=ImRm/(I−Im)R_{sh} = I_mR_m/(I-I_m) extends its current range, and a high series multiplier Rse=V/Im−RmR_{se} = V/I_m - R_m extends its voltage range.

The core idea

How the coil deflects. A PMMC instrument has a light rectangular coil of NN turns and area AA, wound on an aluminium former and suspended in the radial field of a permanent magnet; soft-iron pole pieces shape that field so the flux density BB is uniform wherever the coil sits. When a current II flows through the coil, each side experiences a force F=BILF = BIL, and the pair of forces on opposite sides of the coil combine into a deflecting torque proportional to the current:

Td=NBAI=GIT_d = NBAI = GI

where: NN is the number of turns, BB is the (uniform, radial) flux density in tesla, AA is the coil area in square metres, II is the coil current in amperes, and G=NBAG = NBA bundles the fixed properties of the instrument into a single constant.

Why the scale is uniform. The deflecting torque is opposed by a controlling torque from hairsprings (or a taut-band suspension), which grows in direct proportion to the deflection angle θ\theta: Tc=KθT_c = K\theta, with KK the spring constant. The pointer settles wherever the two torques balance, Td=TcT_d = T_c, giving

θ=GK I\theta = \frac{G}{K}\,I

Deflection is therefore directly proportional to current, which is exactly why PMMC scales are evenly spaced — unlike, say, a hot-wire meter's cramped, non-uniform scale.

Why PMMC reads DC only. Because Td=GIT_d = GI changes sign whenever II reverses, an alternating current produces a torque that alternates at the supply frequency. The coil's own inertia cannot follow such rapid reversals, so the average torque over a full cycle is zero and the pointer simply sits near zero — a PMMC movement gives a genuine, steady deflection only for direct current.

Figure
A PMMC coil shown between the poles of a permanent magnet with the shunt resistor drawn in parallel across the coil and the multiplier resistor drawn in series with the coil
PMMC coil, shunt (parallel) and multiplier (series) coil between magnet poles N S coil pointer swings, T_d = NBAI ammeter (shunt) and voltmeter (multiplier) G shunt R_sh parallel, low R R_se multiplier, series, high R Shunt bypasses excess current; multiplier drops excess voltage.
Derivation ADesigning the shunt and multiplier resistors

The goal is to turn a fragile PMMC coil that can only tolerate a full-scale current ImI_m into an ammeter reading a much larger current II, and separately into a voltmeter reading a much larger voltage VV, by choosing exactly one extra resistor in each case, using nothing beyond Ohm's law and Kirchhoff's laws applied to the coil-plus-resistor combination.

Step 1. Set up the shunt circuit. The coil, of resistance RmR_m, is placed in parallel with a shunt resistor RshR_{sh}, and the combination carries the full line current II. By KCL, the current splits: ImI_m flows through the coil and the remainder, I−ImI - I_m, flows through the shunt.

Step 2. Apply Ohm's law to both branches. Because the coil and the shunt are in parallel, they share the same voltage across their terminals: Vparallel=ImRmV_{\text{parallel}} = I_m R_m (across the coil) and also Vparallel=(I−Im)RshV_{\text{parallel}} = (I - I_m)R_{sh} (across the shunt). Setting these equal gives ImRm=(I−Im)RshI_m R_m = (I - I_m)R_{sh}.

Step 3. Solve for the shunt resistance. Rearranging, Rsh=ImRmI−ImR_{sh} = \dfrac{I_m R_m}{I - I_m}. Since I≫ImI \gg I_m for any useful ammeter range, the denominator is close to II, and RshR_{sh} comes out very small compared with RmR_m — the shunt is a low-resistance, heavy-gauge element precisely because it must carry nearly all of the line current.

Step 4. Now set up the multiplier circuit. Here the coil and a series resistor RseR_{se} share the same current, ImI_m, because they are in series across the full voltage VV to be measured. By KVL, the applied voltage splits across the two: V=ImRm+ImRse=Im(Rm+Rse)V = I_m R_m + I_m R_{se} = I_m(R_m + R_{se}).

Step 5. Solve for the multiplier resistance. Rearranging, Rse=VIm−RmR_{se} = \dfrac{V}{I_m} - R_m. Because ImI_m is tiny (milliamperes) while VV can be tens or hundreds of volts, RseR_{se} comes out very large compared with RmR_m — the multiplier is a high-resistance element so that it, not the coil, drops nearly all of the measured voltage.

Worked example

A PMMC galvanometer has coil resistance Rm=5 ΩR_m = 5\ \Omega and gives full-scale deflection for Im=1 mAI_m = 1\ \text{mA}. Design (a) a shunt to convert it into a 0 to 1 A ammeter, and (b) a multiplier to convert the same coil into a 0 to 150 V voltmeter.

Step 1. Shunt for the 1 A range. The full-scale ammeter current is I=1 A=1000 mAI = 1\ \text{A} = 1000\ \text{mA}, so the shunt must carry I−Im=999 mAI - I_m = 999\ \text{mA} while the coil carries 1 mA1\ \text{mA}. Using Rsh=ImRm/(I−Im)R_{sh} = I_mR_m/(I - I_m):

Rsh=(1×10−3)(5)1−1×10−3=5×10−30.999≈5.005×10−3 ΩR_{sh} = \frac{(1\times10^{-3})(5)}{1 - 1\times10^{-3}} = \frac{5\times10^{-3}}{0.999} \approx 5.005\times10^{-3}\ \Omega

So a shunt of about 5.0 mΩ5.0\ \text{m}\Omega in parallel with the coil diverts 999 mA999\ \text{mA} around it while 1 mA1\ \text{mA} flows through the coil, giving full-scale deflection at a total line current of 1 A. Notice how small RshR_{sh} is next to RmR_m, confirming that shunts must be low-resistance.

Step 2. Multiplier for the 150 V range. Using Rse=V/Im−RmR_{se} = V/I_m - R_m:

Rse=1501×10−3−5=150,000−5=149,995 Ω≈150 kΩR_{se} = \frac{150}{1\times10^{-3}} - 5 = 150{,}000 - 5 = 149{,}995\ \Omega \approx 150\ \text{k}\Omega

The total voltmeter resistance is Rm+Rse=150 kΩR_m + R_{se} = 150\ \text{k}\Omega. Check: at full scale, Im=V/(Rm+Rse)=150/150,000=1 mAI_m = V/(R_m + R_{se}) = 150/150{,}000 = 1\ \text{mA}, matching the coil's rated current — the design is consistent.

Step 3. Read off the sensitivity. The voltmeter's "ohms-per-volt" figure is 150,000/150=1000 Ω/V150{,}000/150 = 1000\ \Omega/\text{V}, which is simply 1/Im1/I_m. A smaller ImI_m always gives a more sensitive (higher-resistance) voltmeter, which disturbs the circuit being measured less. These two results, Rsh≈5.0 mΩR_{sh} \approx 5.0\ \text{m}\Omega for the ammeter and Rse≈150 kΩR_{se} \approx 150\ \text{k}\Omega for the voltmeter, are the final designed values for this galvanometer.

Common traps

  • Writing the shunt denominator as just II instead of I−ImI - I_m; the shunt only carries the difference current, since the coil itself still takes ImI_m.
  • Assuming a PMMC ammeter or voltmeter can read AC directly; without an added rectifier stage (out of scope here), the average torque over a cycle is zero and the pointer will not deflect meaningfully.
  • Swapping shunt and multiplier placement: shunt is always parallel (low resistance, for current), multiplier is always series (high resistance, for voltage). Reversing them can destroy the range and possibly the coil.
  • Forgetting to subtract RmR_m when computing Rse=V/Im−RmR_{se} = V/I_m - R_m — the coil is already part of the series path, so its resistance must come out of the total, not be added on top.

Ask the tutor: Want to try designing a shunt or multiplier for a galvanometer with your own RmR_m and ImI_m values, or see what happens if you accidentally swap the shunt and multiplier connections?

Why does the shunt resistance come out so much smaller than the coil resistance?

Because the shunt must carry almost the entire line current while the coil carries only its tiny rated current ImI_m. Since the two are in parallel and share the same voltage, a much larger current through the shunt requires a proportionally much smaller resistance, from Rsh=ImRm/(I−Im)R_{sh} = I_mR_m/(I-I_m). The bigger the ratio of the ammeter's full-scale current to ImI_m, the smaller RshR_{sh} must be relative to RmR_m.

Why is a more sensitive voltmeter one with a smaller Im?

Voltmeter sensitivity is quoted as ohms-per-volt, which works out to exactly 1/Im1/I_m. A smaller ImI_m means the coil needs a larger multiplier resistance to still draw only ImI_m at full scale, so the total voltmeter resistance is higher. A higher-resistance voltmeter draws less current from the circuit under test, disturbing it less and giving a reading closer to the true, unloaded voltage.

Can the same shunt formula be used if the galvanometer's Im is given in microamperes instead of milliamperes?

Yes, the formula Rsh=ImRm/(I−Im)R_{sh} = I_mR_m/(I-I_m) is unit-independent as long as ImI_m and II are expressed in the same units before you subtract or divide. The usual slip is mixing microamperes and amperes in the same calculation; convert both currents to one consistent unit first, then apply the formula exactly as shown in the worked example.